J Flores1, C Lancha, E Pérez Rodríguez, A García Avello, E Bollo, L J García Frade.
Abstract
Three different assays for fibrin/fibrinogen degradation products (FDP) were evaluated in patients with suspected pulmonary embolism (PE) as rapid screening tests with the aim of evaluating whether they could be used either as a substitute of ventilation/perfusion lung scanning or to supplement scintigraphy in patients in whom the scan was inconclusive (low or intermediate probability). D-Dimer by enzyme-linked immunosorbent assay (ELISA) and latex and total FDP by ELISA were measured prospectively in 85 consecutive patients with suspected PE. With a cutoff of 500 ng/ml, D-dimer by ELISA presented a 96% sensitivity and a 42% specificity, with positive and negative predictive values of 49 and 96%, respectively. D-Dimer by latex, also with a cutoff of 500 ng/ml showed a 93% sensitivity and 29% specificity, with positive and negative predictive values of 43 and 89%. For total FDP, with a cutoff of 900 ng/ml, the sensitivity and specificity were 96 and 26% respectively, with positive and negative predictive values of 42 and 93%. A normal assay may have reduced the necessity of a ventilation/perfusion only in 28% patients with D-dimer ELISA, 21% with D-dimer latex and 17% with total FDP ELISA and with a possible error of 4, 11 and 7%, respectively.(ABSTRACT TRUNCATED AT 250 WORDS)Entities:
Mesh:
Substances:
Year: 1995 PMID: 8560092 DOI: 10.1159/000196459
Source DB: PubMed Journal: Respiration ISSN: 0025-7931 Impact factor: 3.580