Literature DB >> 8557132

Manual versus computer-automated semen analyses. Part I. Comparison of counting chambers.

J E Johnson1, W R Boone, D W Blackhurst.   

Abstract

OBJECTIVE: To determine the accuracy and precision of counting chambers analyzed manually and with a computer-automated semen analyzer (CASA; Hamilton-Thorne Research, Beverly, MA).
DESIGN: Prospective study using comparative measurements of sperm concentration, motility and concentration of latex beads with three types of counting chambers: hemacytometers, 12 and 20 MicroCell (Conception Technologies, Inc., La Jolla, CA) Chambers, and Makler (Sefi-Medical Instruments, Haifa, Israel) Chambers.
SETTING: A hospital-based Andrology laboratory. PATIENTS: Male partners of couples undergoing infertility evaluation. MAIN OUTCOME MEASURES: Experiment I: measurements of sperm concentration were evaluated within hemacytometers; Experiment II: measurements of sperm concentration in the 35 to 50 x 10(6)/mL range and sperm motility were determined using a known concentration of beads.
RESULTS: Experiment I: differences were demonstrated within two of eight hemacytometers (side 1 versus side 2): however, no significant effect of hemacytometer on variation in sperm concentration measurements was observed. Experiment II: CASA-analyzed 20 MicroCell Chambers demonstrated the best precision for sperm concentration (intraclass correlation coefficient = 0.93) and motility (intraclass correlation coefficient = 0.88). Experiment III: 20 MicroCell Chambers most accurately determined known bead concentration (35 +/- 5 x 10(6)/mL) whether analyzed on CASA (34.9 x 10(6)/mL) or manually (35.2 x 10(6)/mL).
CONCLUSIONS: 20 MicroCell Chambers proved to be accurate and precise for determining concentration and motility of semen specimens whether analyzed manually or with CASA.

Entities:  

Mesh:

Year:  1996        PMID: 8557132     DOI: 10.1016/s0015-0282(16)58043-1

Source DB:  PubMed          Journal:  Fertil Steril        ISSN: 0015-0282            Impact factor:   7.329


  8 in total

1.  Extending the coincubation time of gametes improves in vitro fertilization.

Authors:  W R Boone; J E Johnson
Journal:  J Assist Reprod Genet       Date:  2001-01       Impact factor: 3.412

2.  Can Westgard Quality Control Rules determine the suitability of frozen sperm pellets as a control material for computer assisted semen analyzers?

Authors:  Jane E Johnson; Dawn W Blackhurt; William R Boone
Journal:  J Assist Reprod Genet       Date:  2003-01       Impact factor: 3.412

3.  Does seasonality alter intrauterine insemination outcomes: a 5-year study.

Authors:  J Glenn Proctor; Dawn W Blackhurst; William R Boone
Journal:  J Assist Reprod Genet       Date:  2004-07       Impact factor: 3.412

4.  Accu-Beads as a quality control measure for manual and automated methods of measuring sperm concentration-an observational study.

Authors:  Darian D Vernon; Jane E Johnson; Angela M Houwing; H Lee Higdon; William R Boone
Journal:  J Assist Reprod Genet       Date:  2013-10-31       Impact factor: 3.412

5.  Comparative study of Sperm Motility Analysis System and conventional microscopic semen analysis.

Authors:  Kazuhiko Komori; Akira Tsujimura; Sumio Ishijima; Phanu Tanjapatkul; Kazutoshi Fujita; Yasuhiro Matsuoka; Tetsuya Takao; Yasushi Miyagawa; Shingo Takada; Akihiko Okuyama
Journal:  Reprod Med Biol       Date:  2006-08-09

6.  Double-blind prospective study comparing two automated sperm analyzers versus manual semen assessment.

Authors:  J Lammers; C Splingart; P Barrière; M Jean; T Fréour
Journal:  J Assist Reprod Genet       Date:  2013-11-16       Impact factor: 3.412

Review 7.  Current updates on laboratory techniques for the diagnosis of male reproductive failure.

Authors:  Suresh C Sikka; Wayne J G Hellstrom
Journal:  Asian J Androl       Date:  2016 May-Jun       Impact factor: 3.285

8.  Accuracy Evaluation of The Depth of Six Kinds of Sperm Counting Chambers for both Manual and Computer-Aided Semen Analyses.

Authors:  Jin-Chun Lu; Ru-Qian Yue; Rui-Xiang Feng; Ling-Zhu Kong; Yuan-Cheng Xu
Journal:  Int J Fertil Steril       Date:  2015-12-23
  8 in total