| Literature DB >> 26985341 |
Jin-Chun Lu1, Ru-Qian Yue2, Rui-Xiang Feng1, Ling-Zhu Kong2, Yuan-Cheng Xu2.
Abstract
BACKGROUND: Although the depth of the counting chamber is an important factor influencing sperm counting, no research has yet been reported on the measurement and comparison of the depth of the chamber. We measured the exact depths of six kinds of sperm counting chambers and evaluated their accuracy.Entities:
Keywords: Depth; Measurement; Sperm Counting Chamber
Year: 2015 PMID: 26985341 PMCID: PMC4793174 DOI: 10.22074/ijfs.2015.4612
Source DB: PubMed Journal: Int J Fertil Steril ISSN: 2008-0778
Determination results of six kinds of sperm counting chambers’ depth
| Sperm counting chamber | n | Depth of chamber(range) (μm) | CV between chambers (%) | Variation of depth between 5 points (Max-Min, μm) | CV between five points (%) | Acceptability(%) |
|---|---|---|---|---|---|---|
| Macro | 32 | 10.19 ± 0.48 | 4.71 | 0.14 | 0.58 | 65.63 |
| (9.25-10.95) | (0.01-0.37) | (0.06-2.55) | ||||
| Makler (New) | 5 | 11.07 ± 0.41 | 3.70 | 0.086 | 0.30 | 20 |
| (10.50-11.60) | (0.04-0.14) | (0.15-0.60) | ||||
| Makler (Be used) | 19 | 12.72 ± 1.08 | 8.49 | 0.43 | 1.33 | 0 |
| (11.32-14.64) | (0.03-1.27) | (0.09-3.88) | ||||
| Cell-VU | 20 | 24.22 ± 2.58 | 10.65 | 0 | ||
| (18.21-26.83) | ||||||
| Left | 20 | 23.94 ± 2.96 | 12.36 | 1.64 | 2.66 | 5 |
| (17.01-28.34) | (0.35-3.12) | (0.66-4.67) | ||||
| Right | 20 | 24.49 ± 2.55 | 10.41 | 2.21 | 3.47 | 5 |
| (18.68-27.83) | (0.58-4.09) | (0.84-6.33) | ||||
| Geoffrey | 34 | 10.00 ± 0.28 | 2.80 | 94.12 | ||
| (9.54-10.62) | ||||||
| Left | 34 | 10.01 ± 0.29 | 2.90 | 0.11 | 0.45 | 100 |
| (9.54-10.62) | (0.02-0.21) | (0.11-0.85) | ||||
| Right | 34 | 9.99 ± 0.28 | 2.80 | 0.11 | 0.47 | 94.12 |
| (9.56-10.57) | (0.01-0.36) | (0.05-1.61) | ||||
| GoldCyto | 20 | 23.76 ± 2.15 | 9.06 | 0 | ||
| (18.43-26.63) | ||||||
| A | 20 | 22.19 ± 2.62 | 11.81 | 1.81 | 3.27 | 30 |
| (17.68-27.68) | (0.39-4.13) | (0.88-5.59) | ||||
| B | 20 | 24.79 ± 2.52 | 10.17 | 1.99 | 3.23 | 5 |
| (19.30-29.31)* | (0.24-4.80) | (0.43-7.52) | ||||
| C | 20 | 24.96 ± 3.10 | 12.42 | 1.79 | 2.96 | 5 |
| (18.79-29.67)* | (0.37-3.90) | (0.69-5.96) | ||||
| D | 20 | 23.11 ± 2.89 | 12.51 | 1.93 | 3.22 | 10 |
| (17.97-27.81) | (0.15-4.29) | (0.24-6.67) | ||||
| Leja | 20.49 ± 0.22 | 1.06 | 35 | |||
| (20.27-21.34) | ||||||
| A | 20 | 20.99 ± 0.39 | 1.86 | 0.45 | 0.85 | 65 |
| (20.53-21.71) | (0.13-0.80) | (0.25-1.67) | ||||
| B | 20 | 20.37 ± 0.58 | 2.85 | 0.17 | 0.33 | 95 |
| (19.99-22.68)# | (0.05-0.42) | (0.10-0.85) | ||||
| C | 20 | 20.30 ± 0.34 | 1.67 | 0.15 | 0.29 | 95 |
| (19.99-21.67)# | (0.03-0.44) | (0.05-0.87) | ||||
| D | 20 | 20.34 ± 0.14 | 0.69 | 0.18 | 0.35 | 100 |
| (20.15-20.65)# | (0.04-0.42) | (0.07-0.96) | ||||
| E | 20 | 20.88 ± 0.42 | 2.01 | 0.31 | 0.59 | 65 |
| (20.28-21.64) | (0.11-0.60) | (0.21-1.29) | ||||
| F | 20 | 20.31 ± 0.26 | 1.28 | 0.21 | 0.44 | 100 |
| (20.00-20.92)# | (0.07-0.38) | (0.14-0.86) | ||||
| G | 20 | 20.17 ± 0.46 | 2.28 | 0.11 | 0.22 | 95 |
| (19.89-22.07)# | (0.03-0.4) | (0.04-0.69) | ||||
| H | 20 | 20.55 ± 0.30 | 1.46 | 0.33 | 0.63 | 95 |
| (20.23-21.49)#∆ | (0.04-0.61) | (0.08-1.14) | ||||
For Cell-VU and Geoffrey chambers, the difference between the left and right chambers was compared with paired t test. There was no significant difference in depth between the left and right chambers of Cell-VU chambers (t=-1.231, P=0.233) and between the left and right chambers of Geoffrey chambers (t=0.801, P=0.429). The comparison of new and used Makler chambers was analyzed with independent-samples t test, and there was significant difference between them (t=5.325, P<0.001). For GoldCyto and Leja chambers, the difference between the chambers was compared with LSD-t test. CV; Coefficient of variation *; P<0.05 vs. A and D in the same group, #; P<0.05 vs. A and E in the same group and Δ; P<0.05 vs. G in the same group. There was no significant difference between other chambers in the same group (P>0.05).