Literature DB >> 8555573

Small intrapulmonary artery lung prototypes: design, construction, and in vitro water testing.

M T Snider1, K M High, R B Richard, G Panol, E A Campbell, C V Service, J K Stene, J S Ultman.   

Abstract

Blind-ended, hollow fibers mounted on a pulmonary artery catheter may allow O2 and CO2 transfer in the vena cava, right ventricle, and pulmonary artery. The effects of fiber length, manifold number, and gas oscillation on mass and momentum transfer with water perfusate using mass spectrometry and mass flow controllers were studied. Manifolds with 112-196 microporous polypropylene fibers were mounted on 8 Fr multiple lumen, commercially available pulmonary artery catheters. Fiber lengths varied from 0.5 to 16 cm and surface areas from 7 to 220 cm2. Prototypes with 2 cm long fibers were constructed with 1-15 manifolds. A two manifold prototype with 8 cm long fibers and a surface area of 378 cm2 was also studied. The transfer failed to scale with manifold number because the steady gas flow was maldistributed to the manifolds. Oscillating gas pressures from 780 to 76 mmHg absolute at a rate of 40 cycles/min increased CO2 transfer up to 15-fold and O2 transfer up to 2.5-fold. Oscillation also corrected the maldistribution. Optimal fiber lengths of 3 and 1 cm for O2 and CO2, respectively, were seen with steady gas flow, and 8 cm for both with oscillatory gas flow.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8555573     DOI: 10.1097/00002480-199407000-00057

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  3 in total

1.  Effect of impeller design and spacing on gas exchange in a percutaneous respiratory assist catheter.

Authors:  R Garrett Jeffries; Brian J Frankowski; Greg W Burgreen; William J Federspiel
Journal:  Artif Organs       Date:  2014-04-22       Impact factor: 3.094

2.  Evaluation of fiber bundle rotation for enhancing gas exchange in a respiratory assist catheter.

Authors:  Heide J Eash; Kevin M Mihelc; Brian J Frankowski; Brack G Hattler; William J Federspiel
Journal:  ASAIO J       Date:  2007 May-Jun       Impact factor: 2.872

3.  Development of a novel intravascular oxygenator catheter: Oxygen mass transfer properties across nonporous hollow fiber membranes.

Authors:  Stewart Farling; Tobias L Straube; Travis P Vesel; Nick Bottenus; Bruce Klitzman; Ira M Cheifetz; Marc A Deshusses
Journal:  Biotechnol Bioeng       Date:  2020-10-07       Impact factor: 4.530