Literature DB >> 8550334

Determination of excimer laser ablation rate of the human cornea using in vivo Scheimpflug videography.

H J Huebscher1, U Genth, T Seiler.   

Abstract

PURPOSE: To determine in vivo the amount of human corneal tissue removed by each excimer laser pulse, the so-called ablation rate, during photorefractive keratectomy (PRK). There is confusion in the literature because the experimentally determined ablation rate of 0.4 to 0.5 microns per pulse differs from the nominal ablation rate of 0.23 to 0.3 microns per pulse, which is the value used in clinical procedures.
METHODS: Eleven eyes of 11 patients were treated with PRK for correction of myopia. The corneal curvature was determined by Scheimpflug videography before and immediately after surgery. Starting from this curvature change, the authors calculated the real ablation rate.
RESULTS: The real ablation rate is coincident with the nominal ablation rate and differs significantly from the ablation rate derived from deep keratectomy experiments.
CONCLUSIONS: The outer layers of the cornea show significantly different ablation behavior than the deeper stroma. This information has clinical relevance for the predictability of intrastromal excimer laser procedures.

Entities:  

Mesh:

Year:  1996        PMID: 8550334

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  5 in total

1.  [Online optical coherence pachymetry in laser in situ keratomileusis].

Authors:  C Wirbelauer; H Häberle; D T Pham
Journal:  Ophthalmologe       Date:  2004-02       Impact factor: 1.059

2.  Experimental evaluation of online optical coherence pachymetry for corneal refractive surgery.

Authors:  Christopher Wirbelauer; Henning Aurich; Jan Jaroszewski; Christian Hartmann; Duy Thoai Pham
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-11-14       Impact factor: 3.117

3.  Analytical optimization of the ablation efficiency at normal and non-normal incidence for generic super Gaussian beam profiles.

Authors:  Samuel Arba-Mosquera; Shwetabh Verma
Journal:  Biomed Opt Express       Date:  2013-07-24       Impact factor: 3.732

4.  Evaluation of the efficacy of excimer laser ablation of cross-linked porcine cornea.

Authors:  Shihao Chen; Yini Li; Aleksander Stojanovic; Jia Zhang; Yibo Wang; Qinmei Wang; Theo Seiler
Journal:  PLoS One       Date:  2012-10-03       Impact factor: 3.240

5.  Online optical coherence pachymetry to evaluate intraoperative ablation parameters in LASIK.

Authors:  Christopher Wirbelauer; Henning Aurich; Duy Thoai Pham
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-11-22       Impact factor: 3.535

  5 in total