Literature DB >> 8543210

Corneal endothelial permeability and aqueous humor flow using a standard protocol.

J A van Best1, M Diestelhorst, E Leite, S Fantaguzzi, R Schalnus.   

Abstract

BACKGROUND: The study was carried out to compare corneal endothelial permeability and aqueous flow values of healthy volunteers measured in different countries with identical fluorophotometers using a standardized protocol.
METHOD: Healthy volunteers aged between 20 and 70 years were studied in five European cities. Fluorescence scans of the anterior segment of both eyes were made using a commercial fluorophotometer. Beginning 4 h after instillation of four drops of fluorescein 10%, 12 scans of the anterior segment of each eye were performed in 2 h. The values of corneal endothelial permeability and aqueous flow were calculated with standardized software from the decay of the fluorescein concentration in the cornea and anterior chamber.
RESULTS: The mean permeability values (x 10(-4) cm.min-1) +/- SD were 3.7 +/- 1.6 (n = 19; Coimbra, Portugal), 4.3 +/- 1.1 (n = 19; Frankfurt, Germany), 3.9 +/- 0.9 (n = 19; Leiden, The Netherlands) and 5.4 +/- 1.2 (n = 10; Milan, Italy). The values were not significantly different (ANOVA, P > 0.3), except those in Milan. The mean flow values (microliters.min-1) +/- SD were 2.3 +/- 0.9 (n = 17; Coimbra), 1.9 +/- 0.7 (n = 10; Cologne, Germany), 2.6 +/- 1.2 (n = 19; Frankfurt), 2.0 +/- 0.6 (n = 19; Leiden) and 1.7 +/- 0.8 (n = 10; Milan). The values were not significantly different (Kruskal-Wallis test, P > 0.1).
CONCLUSIONS: Permeability and flow values in the different cities had similar values and standard deviations. The Concerted Action demonstrated the usefulness of a standardized protocol.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8543210     DOI: 10.1007/bf00404710

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  23 in total

1.  Effect of long-term contact lens wear on corneal endothelial cell morphology and function.

Authors:  K H Carlson; W M Bourne; R F Brubaker
Journal:  Invest Ophthalmol Vis Sci       Date:  1988-02       Impact factor: 4.799

2.  Fluorophotometry and the rate of aqueous flow in man. I. Instrumentation and normal values.

Authors:  J N Bloom; R Z Levene; G Thomas; R Kimura
Journal:  Arch Ophthalmol       Date:  1976-03

3.  Blood aqueous barrier permeability versus age by fluorophotometry.

Authors:  J A van Best; J P Kappelhof; L Laterveer; J A Oosterhuis
Journal:  Curr Eye Res       Date:  1987-07       Impact factor: 2.424

4.  [Fluorophotometry by instillation. I. Flow of the aqueous humor and corneal endothelial permeability].

Authors:  L M Coulangeon; J M Menerath; P Sole
Journal:  J Fr Ophtalmol       Date:  1987       Impact factor: 0.818

5.  Kinetics of intraocular penetration of topical fluorescein: analysis by new method.

Authors:  M Araie
Journal:  Jpn J Ophthalmol       Date:  1983       Impact factor: 2.447

6.  Measurement of aqueous humor flow with scanning ocular fluorophotometers.

Authors:  J E Topper; J McLaren; R F Brubaker
Journal:  Curr Eye Res       Date:  1984-12       Impact factor: 2.424

7.  Endothelial barrier function after phacoemulsification: a comparison between diabetic and non-diabetic patients.

Authors:  M Goebbels; M Spitznas
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1991       Impact factor: 3.117

Review 8.  Flow of aqueous humor in humans [The Friedenwald Lecture].

Authors:  R F Brubaker
Journal:  Invest Ophthalmol Vis Sci       Date:  1991-12       Impact factor: 4.799

9.  The effect of age on aqueous humor formation in man.

Authors:  R F Brubaker; S Nagataki; D J Townsend; R R Burns; R G Higgins; W Wentworth
Journal:  Ophthalmology       Date:  1981-03       Impact factor: 12.079

View more
  2 in total

1.  Diffusion coefficient through the blood-aqueous barrier using a standard protocol.

Authors:  J van Best; J B del Castillo; M Diestelhorst; B Heintz; E Leite; L F Liesenborghs; R Schalnus
Journal:  Br J Ophthalmol       Date:  1996-04       Impact factor: 4.638

2.  Corneal epithelial permeability to fluorescein in humans by a multi-drop method.

Authors:  Sangly P Srinivas; Arushi Goyal; Deepti P Talele; Sanjay Mahadik; Rachapalle Reddi Sudhir; P Pavani Murthy; Sudhir Ranganath; Uday B Kompella; Prema Padmanabhan
Journal:  PLoS One       Date:  2018-06-19       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.