Visual display format affects the ability of anesthesiologists to detect acute physiologic changes. A laboratory study employing a clinical display simulator.

K Gurushanthaiah1, M B Weinger, C E Englund.   

Abstract

BACKGROUND: Anesthesiologists use data presented on visual displays to monitor patients' physiologic status. Although studies in nonmedical fields have suggested differential effects on performance among display formats, few studies have examined the effect of display format on anesthesiologist monitoring performance.
METHODS: A computer-based clinical display simulator was developed to evaluate the efficacy of three currently used display formats (numeric, histogram, or polygon displays) in a partial-task laboratory simulation. The subjects' task consisted solely of detecting any changes in the values of the physiologic variables depicted on a simulated clinical display. Response latency and accuracy were used as measures of performance.
RESULTS: Thirteen anesthesia residents and five nonmedical volunteers, were enrolled as subjects. Use of either the histogram or polygon displays significantly improved response latencies and allowed greater accuracy compared with the numeric display in the anesthesia residents. Neither response latency nor accuracy improved with additional exposure to these displays. In contrast, display format did not significantly affect response latency or accuracy in the nonmedical volunteers.
CONCLUSIONS: The results of this study suggest that graphic displays may enhance the detection of acute changes in patient physiologic status during anesthesia administration. This research also demonstrates the importance of assessing performance on clinical devices by studying actual users rather than random subjects. Further research is required to elucidate the display elements and characteristics that best support different aspects of the anesthesiologist's monitoring tasks.

Entities:  

Mesh:

Year:  1995        PMID: 8533911     DOI: 10.1097/00000542-199512000-00009

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  15 in total

1.  The employment of an iterative design process to develop a pulmonary graphical display.

Authors:  S Blake Wachter; Jim Agutter; Noah Syroid; Frank Drews; Matthew B Weinger; Dwayne Westenskow
Journal:  J Am Med Inform Assoc       Date:  2003-03-28       Impact factor: 4.497

2.  Specific elements of a new hemodynamics display improves the performance of anesthesiologists.

Authors:  G T Blike; S D Surgenor; K Whalen; J Jensen
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

3.  A graphical object display improves anesthesiologists' performance on a simulated diagnostic task.

Authors:  G T Blike; S D Surgenor; K Whalen
Journal:  J Clin Monit Comput       Date:  1999-01       Impact factor: 2.502

4.  The challenges of human engineering research.

Authors:  G T Blike
Journal:  J Clin Monit Comput       Date:  1999-12       Impact factor: 2.502

5.  Usability evaluation of a GUI prototype for a ventilator machine.

Authors:  Yuanhua Liu; Anna-Lisa Osvalder
Journal:  J Clin Monit Comput       Date:  2004-12       Impact factor: 2.502

6.  Technologies and solutions for data display in the operating room.

Authors:  Noemi Bitterman
Journal:  J Clin Monit Comput       Date:  2006-05-11       Impact factor: 2.502

7.  An integrated graphic data display improves detection and identification of critical events during anesthesia.

Authors:  P Michels; D Gravenstein; D R Westenskow
Journal:  J Clin Monit       Date:  1997-07

8.  Human factors research in anesthesia patient safety.

Authors:  M B Weinger; J Slagle
Journal:  Proc AMIA Symp       Date:  2001

9.  The design and evaluation of a graphical display for laboratory data.

Authors:  David T Bauer; Stephanie Guerlain; Patrick J Brown
Journal:  J Am Med Inform Assoc       Date:  2010 Jul-Aug       Impact factor: 4.497

Review 10.  Evaluations of physiological monitoring displays: a systematic review.

Authors:  Matthias Görges; Nancy Staggers
Journal:  J Clin Monit Comput       Date:  2007-12-07       Impact factor: 2.502