Implantable cardioverter-defibrillator therapy in the absence of significant symptoms. Rhythm diagnosis and management aided by stored electrogram analysis.

B G Hook1, D J Callans, R B Kleiman, B T Flores, F E Marchlinski.   

Abstract

BACKGROUND: This report describes the value of stored ventricular electrogram analysis in the diagnosis and management of patients experiencing minimal or no symptoms before implantable cardioverter-defibrillator (ICD) therapy. METHODS AND
RESULTS: The study population included 48 patients who received the Cadence Tiered Therapy Defibrillator System, an investigational third-generation ICD with ventricular electrogram storage capabilities. Criteria for arrhythmia diagnosis were based on analysis of the electrogram rate, RR interval variability, and morphology. Twenty-nine of the 48 patients (60%) experienced at least one episode of antitachycardia pacing or shock (one shock or more in 25 of 29 patients) that was preceded by minimal or no symptoms during a mean follow-up of 15.1 +/- 7.8 months. There were 194 tachycardia episodes registered by the device, including 101 for which ventricular electrograms were stored and available for analysis. Of the 101 stored electrograms, 74 were classified as ventricular tachycardia (VT), 24 as non-VT rhythms (atrial fibrillation, 13; supraventricular tachycardia, six; rate-sensing lead disruption, four; T wave oversensing, one), and only three as indeterminate rhythms. Based on the electrogram analysis, changes in tachycardia detection criteria and/or antiarrhythmic drug regimens were implemented and were associated with a reduction in the number of device responses for non-VT rhythms from 24 during the initial study period to three during 11.0 +/- 7.2 months of additional follow-up.
CONCLUSIONS: ICD responses in the absence of symptoms are relatively common in third-generation devices with antitachycardia pacing capabilities. Despite potential limitations such as the effect of bundle branch block on the electrogram morphology during supraventricular tachycardia, the availability of electrogram storage capabilities allowed a presumptive diagnosis of the events precipitating asymptomatic device responses. Device reprogramming based on analysis of stored electrograms was associated with a dramatic reduction in the incidence of ICD responses for non-VT rhythms.

Entities:  

Mesh:

Year:  1993        PMID: 8504502     DOI: 10.1161/01.cir.87.6.1897

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  9 in total

1.  Prospective evaluation of the quality and long-term stability of atrial signals in non-thoracotomy defibrillation electrodes: comparison of four different endocardial electrograms.

Authors:  C Wolpert; W Jung; S Spehl; B Schumacher; H Omran; B Esmailzadeh; B Lüderitz
Journal:  J Interv Card Electrophysiol       Date:  1998-12       Impact factor: 1.900

2.  Pattern of initiation of monomorphic ventricular tachycardia in recorded intracardiac electrograms.

Authors:  Majid Haghjoo; Arash Arya; Mohammad Ali Sadr-Ameli
Journal:  Indian Pacing Electrophysiol J       Date:  2005-10-01

3.  Case report: inappropriate discharge of an implantable cardioverter-defibrillator caused by the combined count criterion.

Authors:  C Wolpert; W Jung; S Spehl; T Korte; B Lüderitz
Journal:  J Interv Card Electrophysiol       Date:  1998-03       Impact factor: 1.900

4.  [Current status and problems of the entirely subcutaneous ICD (S-ICD®)].

Authors:  F Reinke; A Löher; J Köbe; L Eckardt
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2013-09

5.  Diagnostic value of single versus dual chamber electrograms recorded from an implantable defibrillator.

Authors:  Michael H Kim; David Bruckman; Christian Sticherling; Hakan Oral; Frank Pelosi; Bradley P Knight; Fred Morady; S Adam Strickberger
Journal:  J Interv Card Electrophysiol       Date:  2003-08       Impact factor: 1.900

6.  [Stored electrograms to differenciate between adequate and inadequate ICD therapy].

Authors:  D Müller; E Hoffman; M Fiek; A Grünewald; G Steinbeck
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03

7.  [New algorithms for discrimination between supraventricular and ventricular tachyarrhythmias in patients with implantable cardioverter/defibrillator].

Authors:  J Neuzner; M Schlepper
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03

8.  Benefits of treatment with implantable cardioverter-defibrillators in patients with stable ventricular tachycardia without cardiac arrest.

Authors:  D Böcker; M Block; F Isbruch; C Fastenrath; M Castrucci; D Hammel; H H Scheld; M Borggrefe; G Breithardt
Journal:  Br Heart J       Date:  1995-02

9.  Identifying non-inducible ventricular tachycardia origin utilizing defibrillator electrograms.

Authors:  Cory M Tschabrunn; Elad Anter; Francis E Marchlinski
Journal:  J Interv Card Electrophysiol       Date:  2012-10-27       Impact factor: 1.900