Accuracy and precision of low-dose insulin administration.

S J Casella1, M K Mongilio, L P Plotnick, M P Hesterberg, C A Long.   

Abstract

OBJECTIVE: To determine the lowest dose of concentrated (U100) insulin that can be reproducibly delivered.
METHODS: A telephone survey was used to determine current practices in major pediatric hospitals regarding the administration of low doses of concentrated insulin. A sensitive gravitometric technique was used to determine the error in measurement of low doses of U100 insulin by pediatric nurses and parents of diabetic children.
RESULTS: A telephone survey revealed that doses as low as 0.5 or 1.0 U (5 to 10 microL) are routinely administered in pediatric hospitals. In our study of pediatric nurses, attempts to deliver 0.5, 1.0, and 2.0 U resulted in delivered doses of 0.975 +/- 0.315, 1.638 +/- 0.376, and 2.153 +/- 0.435 U (mean +/- standard deviation of the mean), respectively. The use of 0.3-mL syringes compared to 0.5-mL syringes did not improve accuracy or precision. Taken as a group, parents of children with diabetes were more accurate (mean delivered dose of 1.063 +/- 0.276 for the 1-U dose), but the individual means ranged from 0.641 to 1.300 and coefficients of variation ranged from 5% to 33%. Only three of the seven parents could deliver 1.0 U with acceptable precision and accuracy.
CONCLUSIONS: When currently available syringes are used, insulin injections of less than 20 microL (2 U of U100) have an unacceptably large error. It is recommended that, in the inpatient setting, diluted insulin be used if the prescribed dose is less than 2.0 U.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8502520

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  10 in total

1.  Can variable practice habits and injection port dead-volume put patients at risk?

Authors:  Michael T Kuntz; Roman Dudaryk; Richard R McNeer
Journal:  J Clin Monit Comput       Date:  2018-07-10       Impact factor: 2.502

2.  Novel methodology to determine the accuracy of the OmniPod insulin pump: a key component of the artificial pancreas system.

Authors:  Howard Zisser; Marc Breton; Eyal Dassau; Kateryna Markova; Wendy Bevier; Dale Seborg; Boris Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

3.  Use of Computer Vision to Identify the Frequency and Magnitude of Insulin Syringe Preparation Errors.

Authors:  Ann Cabri; Berit Bagley; Kevin Brown
Journal:  J Diabetes Sci Technol       Date:  2020-08-05

4.  Accuracy and reproducibility of low dose insulin administration using pen-injectors and syringes.

Authors:  M G Gnanalingham; P Newland; C P Smith
Journal:  Arch Dis Child       Date:  1998-07       Impact factor: 3.791

Review 5.  Accuracy of intravenous and enteral preparations involving small volumes for paediatric use: a review.

Authors:  L P Ainscough; J L Ford; C W Morecroft; M Peak; M A Turner; A J Nunn; M Roberts
Journal:  Eur J Hosp Pharm       Date:  2017-01-04

6.  Development of Guidelines for Accurate Measurement of Small Volume Parenteral Products Using Syringes.

Authors:  Melanie A Jordan; Dimpa Choksi; Kelsey Lombard; Lynn R Patton
Journal:  Hosp Pharm       Date:  2019-09-06

7.  Accuracy and Injection Force of the Gla-300 Injection Device Compared With Other Commercialized Disposable Insulin Pens.

Authors:  David Klonoff; Irina Nayberg; Marissa Thonius; Florian See; Mona Abdel-Tawab; Frank Erbstein; Thomas Haak
Journal:  J Diabetes Sci Technol       Date:  2015-08-26

8.  Tolerability, safety and adherence to treatment with insulin detemir injection in the treatment of type 2 diabetes.

Authors:  Athena Philis-Tsimikas
Journal:  Patient Prefer Adherence       Date:  2008-02-02       Impact factor: 2.711

Review 9.  Update on insulin treatment for dogs and cats: insulin dosing pens and more.

Authors:  Ann Thompson; Patty Lathan; Linda Fleeman
Journal:  Vet Med (Auckl)       Date:  2015-04-15

10.  Comparison between a flash glucose monitoring system and a portable blood glucose meter for monitoring dogs with diabetes mellitus.

Authors:  Francesca Del Baldo; Claudia Canton; Silvia Testa; Harry Swales; Ignazio Drudi; Stefania Golinelli; Federico Fracassi
Journal:  J Vet Intern Med       Date:  2020-10-30       Impact factor: 3.333

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.