Evaluation of a computerised polysomnography system.

H Biernacka1, N J Douglas.   

Abstract

BACKGROUND: Manual analysis of sleep, breathing, and oxygenation records is the "gold standard" for diagnosing sleep abnormalities but is time consuming and cumbersome. The accuracy and cost of a computerised sleep analysis system have therefore been investigated.
METHODS: Manual and computerised (CNS Sleep Lab) scores from 43 consecutive clinical sleep studies were prospectively compared for accuracy and the time and costs were recorded.
RESULTS: There were good correlations and no systematic differences between manual and computer scoring for total sleep time, sleep onset latency, and duration of REM sleep. There was a small but clinically insignificant systematic difference in breathing pattern analysis, the number of hypopnoeas/hour being lower with manual than with computer scoring (13 (SE 3) v 15 (SE 3)/hour). There was no difference between computer and manual scoring of the frequency of apnoeas, so the frequency of apnoeas + hypopnoeas was clinically insignificantly higher with computer scoring with a highly significant correlation between the two techniques. The time taken to perform the analyses was not different between the two methods (manual 83 (SE 8) v computer 86 (SE 8) minutes). The computer system was six times more expensive than the manual system and annual running costs, including full maintenance contract and 15% depreciation, were twice as great.
CONCLUSION: The CNS Sleep Lab is sufficiently accurate for use in clinical sleep studies but is significantly more expensive and does not save technician time.

Entities:  

Mesh:

Year:  1993        PMID: 8497829      PMCID: PMC464370          DOI: 10.1136/thx.48.3.280

Source DB:  PubMed          Journal:  Thorax        ISSN: 0040-6376            Impact factor:   9.139


  6 in total

1.  The sleep hypopnea syndrome.

Authors:  G A Gould; K F Whyte; G B Rhind; M A Airlie; J R Catterall; C M Shapiro; N J Douglas
Journal:  Am Rev Respir Dis       Date:  1988-04

2.  Guidelines for the multiple sleep latency test (MSLT): a standard measure of sleepiness.

Authors:  M A Carskadon; W C Dement; M M Mitler; T Roth; P R Westbrook; S Keenan
Journal:  Sleep       Date:  1986-12       Impact factor: 5.849

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  MESAM: a heart rate and snoring recorder for detection of obstructive sleep apnea.

Authors:  T Penzel; G Amend; K Meinzer; J H Peter; P von Wichert
Journal:  Sleep       Date:  1990-04       Impact factor: 5.849

5.  Clinical value of polysomnography.

Authors:  N J Douglas; S Thomas; M A Jan
Journal:  Lancet       Date:  1992-02-08       Impact factor: 79.321

6.  Transient hypoxaemia during sleep in chronic bronchitis and emphysema.

Authors:  N J Douglas; P M Calverley; R J Leggett; H M Brash; D C Flenley; V Brezinova
Journal:  Lancet       Date:  1979-01-06       Impact factor: 79.321

  6 in total
  5 in total

1.  The use of epochs to stage sleep results in incorrect computer-generated AHI values.

Authors:  Mark B Norman; Sally Middleton; Colin E Sullivan
Journal:  Sleep Breath       Date:  2010-04-13       Impact factor: 2.816

2.  Deterministic properties of apnoeas in an abdominal breathing signal.

Authors:  P M Macey; J S Li; R P Ford
Journal:  Med Biol Eng Comput       Date:  1999-05       Impact factor: 2.602

3.  A short-term controlled trial of an adjustable oral appliance for the treatment of mild to moderate obstructive sleep apnoea.

Authors:  K A Ferguson; T Ono; A A Lowe; S al-Majed; L L Love; J A Fleetham
Journal:  Thorax       Date:  1997-04       Impact factor: 9.139

4.  Computerised polysomnography.

Authors:  N J Douglas
Journal:  Thorax       Date:  1994-05       Impact factor: 9.139

5.  Manual vs. automated analysis of polysomnographic recordings in patients with chronic obstructive pulmonary disease.

Authors:  Gerben Stege; Petra J E Vos; P N Richard Dekhuijzen; Pieter H E Hilkens; Marjo J T van de Ven; Yvonne F Heijdra; Frank J J van den Elshout
Journal:  Sleep Breath       Date:  2012-05-10       Impact factor: 2.816

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.