| Literature DB >> 845242 |
J R Stetter, J M Sedlak, K F Blurton.
Abstract
An electrochemical detector for gas chromatographic analysis of hydrogen sulfide is described and its operating characteristics are discussed. The detector operation is based upon the measurement of the current when hydrogen sulfide is electrochemically oxidized at a diffusion electrode. The lower detectable limit was 3 X 10(-12) grams H2S, the precision was 0.5% and analysis was achieved within 2 minutes. Accuracy was limited principally by the preparation of calibration samples and the availability of standard reference gas mixtures.Entities:
Mesh:
Substances:
Year: 1977 PMID: 845242 DOI: 10.1093/chromsci/15.3.125
Source DB: PubMed Journal: J Chromatogr Sci ISSN: 0021-9665 Impact factor: 1.618