Trace hydrogen detection plays an important role in the safety detection of lithium-ion batteries (LIBs) due to the generation and leakage of trace hydrogen in the early stage of LIBs damage. In this work, an amperometric hydrogen sensor based on solid polymer electrolyte was reported. The sandwich device structure was realized, which could directly diffuse the gas from both sides to the three-phase interface (gas/electrode/electrolyte) to participate in the reaction through the optimal design of the gas diffusion path. Then, platinum nanoparticles (Pt-NPs) were loaded on the metal foam by electroplating, and the porous electrode was filled with solid polymer electrolyte. A sensor with high specific surface area, high catalytic activity, and high sensitivity was obtained. Finally, the hydrogen oxidation reaction (HOR) mechanism of the platinum-loaded (Pt-loaded) titanium foam (Ti foam) electrode under both anaerobic and aerobic conditions was verified, and the properties of the sensor was evaluated. The hydrogen sensor with a "sandwich" structure has the advantages of high sensitivity, good stability, low detection limit and low cost, which provides a technical solution for the safety and real-time monitoring of LIBs.
Trace hydrogen detection plays an important role in the safety detection of lithium-ion batteries (LIBs) due to the generation and leakage of trace hydrogen in the early stage of LIBs damage. In this work, an amperometric hydrogen sensor based on solid polymer electrolyte was reported. The sandwich device structure was realized, which could directly diffuse the gas from both sides to the three-phase interface (gas/electrode/electrolyte) to participate in the reaction through the optimal design of the gas diffusion path. Then, platinum nanoparticles (Pt-NPs) were loaded on the metal foam by electroplating, and the porous electrode was filled with solid polymer electrolyte. A sensor with high specific surface area, high catalytic activity, and high sensitivity was obtained. Finally, the hydrogen oxidation reaction (HOR) mechanism of the platinum-loaded (Pt-loaded) titanium foam (Ti foam) electrode under both anaerobic and aerobic conditions was verified, and the properties of the sensor was evaluated. The hydrogen sensor with a "sandwich" structure has the advantages of high sensitivity, good stability, low detection limit and low cost, which provides a technical solution for the safety and real-time monitoring of LIBs.
Recently, the rise
of intelligent wearable electronic devices used
in life, medical treatment, and sports as well as the development
of transportation power systems, aerospace, and other fields have
promoted the wide application of lithium-ion bateries (LIBs).[1,2] However, the damage of LIBs often causes serious safety accidents
in a short time, which means that it is necessary to detect the damage
of LIBs as early as possible.[3−5] At the same time, it is worth
mentioning that trace hydrogen detection is an effective means of
LIBs safety monitoring. The reason is that the thermal runaway caused
by the overcharge of LIBs generate a large number of lithium (Li)
dendrites under high temperature conditions, which react with the
organic electrode solution binder to generate gases such as hydrogen.[6−8] According to the study of Yuan et al., trace hydrogen detection
can advance the alarm time to 639 s before the smoking of the LIBs,
which preliminarily illustrates the feasibility of the scheme.[9]Electrochemical gas sensors, especially
amperometric gas sensors,
are widely used in trace gas detection due to their advantages of
high sensitivity, low manufacturing cost, environmental protection,
low power consumption, and easy integration.[10−13] At the same time, combined with
the international standards of hydrogen detection and the monitoring
requirements of LIBs,[14−17] stricter requirements are put forward for the sensitivity, response
speed, and anti-interference ability of hydrogen detection. Furthermore,
it is necessary to use low-cost hydrogen sensors to rapidly detect
trace hydrogen (ppm level) for the popularization of safety detection
of LIBs.[18,19]The design of the device structure
is one of the key factors that
determine the performance of the amperometric gas sensor. Recently,
screen-printed electrodes (SPEs) based on room temperature ionic liquids
(RTILs) have attracted extensive attention due to their simple structure,
small size, and easy integration.[20] However,
conventional electrode structures based on solid ceramic substrates
have limited gas diffusion rates because gas needs to dissolve and
diffuse in RTIL before reaching the three-phase interface. The difference
is to use a porous ceramic substrate as the gas diffusion barrier
to improve the sensor design as pointed out in our previous study.[21] The gas molecules directly diffuse from the
porous substrate to the three-phase interface without passing through
the electrolyte membrane, which effectively improves the response
speed of the device. Besides, since the interface between the electrolyte
membrane and the electrode is a two-dimensional planar structure,
the area of the effective three-phase interface is small, which limits
the sensitivity of the hydrogen sensors. In contrast, traditional
electrochemical gas sensors based on aqueous solutions of strong acids
or alkalis have limited lifetimes and operating temperature ranges.
However, the electrolyte solution and the porous electrode form the
large three-phase interface, which ensures the high sensitivity of
the sensor. In addition, it is worth mentioning that Korotcenkov et
al. improved the “Clark” type gas sensor.[22] The working electrode (WE) was in direct contact
with most of the electrolyte and in contact with air, and the mass
transfer rate of hydrogen to the three-phase interface was faster.
Then, Blurton et al. adopted fuel cell electrode technology and a
gas diffusion electrode (GDE) for the design of the sensor.[23] Hydrogen diffused through the porous current
collector directly to the three-phase interface, and the sensor can
achieve a fast response, high sensitivity, and low detection limit.
In conclusion, the performance of the sensor can be effectively improved
by optimizing the gas diffusion path and increasing the contact area
between the electrolyte and the electrode, which has reference significance
for the design of the device structure.In this study, a novel
amperometric gas sensor was designed and
fabricated with a metal foam as the electrode, which was filled with
RTIL electrolyte simultaneously. This design enabled a very high specific
surface area and a strong electrochemical activity when loaded with
platinum nanoparticles (Pt-NPs). The sensor can realize the direct
diffusion of gas from both sides to the three-phase interface to participate
in the reaction. At the same time, the electrochemical behavior of
hydrogen gas was explored in combination with a solid substrate sensor
with a planar structure. It is believed that the present work could
provide a fast hydrogen detection for LIBs safety monitoring.
Experimental Section
Reagents and Apparatus
The chemical
reagents and the materials of the experiment used were chloroplatinic
acid (H14Cl6O6Pt, 99.9%, Shanghai
Aladdin Biochemical Technology Co., Ltd.), hydrochloric acid (HCl,
12 M, purity ≥99.7%, Shanghai Aladdin Biochemical Technology
Co., Ltd.), nitric acid (HNO3, 0.1 M, Shanghai Aladdin
Biochemical Technology Co., Ltd.), polyvinylidene fluoride resin (PVDF,
molecular weight 1.1 × 106, Arkema, France), N,N-dimethylformamide (DMF, purity ≥99.8%,
Shanghai Aladdin Biochemical Technology Co., Ltd.), titanium foam
(Ti foam, purity ≥99 %, Yiminglong Electronic Materials Co.,
Ltd. ). The types of RTIL used was 1-butyl-1-methylpyrrolidinium bis(trifluoromethylsulfonyl)imide
([Bmpy][NTf2], purity ≥99 %, Bailingwei Reagent
Co., Ltd.). All experiments were performed at 298 ± 2 K without
special instructions. The analytical gas followed the certified gas
standards (Air Chemistry Co., Ltd., Beijing). The gas concentration
required for the experiment is diluted with standard gas with nitrogen
(purity ≥99.999%) or synthetic air (purity ≥99.999%)
by mass flow controllers (MFC), where the setup for the characterization
of sensor performance is shown in Figure S1. All of electrochemical measurements were carried out with the Electrochemical
Workstation (Netherlands IVIUM Technologies B.V.). Field-emission-scanning
electron microscopy and energy dispersive X-ray detection (FE-SEM
and EDX, Zeiss Gemini Sigma 300), inductively coupled plasma emission
spectrometry (ICP, Agilent ICPOES730), X-ray photoelectron spectroscopy
(XPS, Thermo Escalab 250Xi), and X-ray diffraction (XRD, Bruker D8
Advance) was used to analyze the appearance and microstructure of
the metal foam after platinum loading.
Electroplating
Processes
Taking titanium
foam (Ti foam) as an example, the experimental process of WE loaded
by Pt was briefly described below. Ti foam (3.3 × 10 × 0.5
mm3) was immersed in acetone and then sonicated for 10
min. Then, Ti foam was soaked in hydrochloric acid to remove the oxide
film on the metal surface, according to Figure a. After pretreatment, Ti foam was stored
in alcohol to prevent air oxidation. Next, the chloroplatinic acid
particles were dissolved in a dilute nitric acid solution to obtain
a plating solution with a pH of about 1. The type of acid solution
used (inset in Figure S2b) and the selection
of electroplating potential (Figure S2a) are detailed in the Supporting Information. The electrodes were
cleaned with alcohol after electroplating at 343 K by a potentiostat
(Figure b). The loading
of Pt on Ti foam was controlled according to the amount of electricity
transferred during the electroplating process.
Figure 1
Process of Ti foam (a)
pretreatment and (b) electroplating and
(c) the illustration of the structure of the sensor.
Process of Ti foam (a)
pretreatment and (b) electroplating and
(c) the illustration of the structure of the sensor.
Fabrication of Sensor Devices
Using
[Bmpy][NTf2], PVDF, and DMF as raw materials, the solidified
liquid was prepared by melting and stirring with organic solvent.
The plated and dried Ti foam was placed in a petri dish and coated
twice with the solidified liquid. The porous WE filled with solid
polymer electrolyte was obtained after complete drying in an oven
at 353 K and sintering at 473 K. CE and RE were made by screen-printing
with a porous ceramic substrate. The 3D-printed package was designed
for the device tightened Ti foam electrode and SPE, as shown in Figure c. At the same time,
a solid polymer electrolyte film was spaced between the two layers
of electrodes to prevent the solid polymer electrolyte and Ti foam
from being completely coated, resulting in a short circuit between
WE and RE. A more detailed sensor fabrication process is described
in the Supporting Information.
Results
and Discussion
Material Characterization
Ti foam
was treated with nitric acid to remove the oxide film on the surface.
However, it was found that the surface of Ti foam was slightly corroded,
compared with Ti foam without pretreatment as shown in Figure a,b. Meanwhile, Ti foam can
be found with a pore size range from 70 to 100 μm. The specific
surface area of Ti foam was increased by acid corrosion, which was
beneficial for Pt plating. The size of the Pt-NPs was about 200 nm,
which were evenly distributed on the surface of Ti foam without obvious
agglomerations. As can be seen from Figure c, no continuous Pt film structure was formed.
Therefore, the high specific surface area of Pt-NPs provided a large
number of electrochemically active sites for the subsequent hydrogen
oxidation reaction (HOR).
Figure 2
SEM images of Ti foam (a) before and (b) after
pretreatment and
(c) after electroplating.
SEM images of Ti foam (a) before and (b) after
pretreatment and
(c) after electroplating.The Pt/Ti foam electrode was consistent with the pattern of Pt
in the orthorhombic state (PDF no. 44-0003) and crystal structure
and slightly different from the Pt in the metallic phase (PDF no.
04-0802), according to XRD patterns in Figure a. However, the XRD pattern indicated the
possible presence of Pt in a combined state. Therefore, the chemical
state of Pt on the Ti foam electrode was further investigated by XPS
analysis. By peaks fitting to the data, the Pt2+ 4f7/2 peak at 72.4 eV and the Pt2+ 4f5/2 peak at 75.6 eV indicated the presence of PtO, and the Pt4+ 4f7/2 peak at 73.5 eV and the Pt4+ 4f5/2 peak at 76.7 eV indicated the presence of PtO2, as shown in Figure b. It can be seen that most of the Pt present on Ti foam electrode
was divalent and tetravalent. It was found that the RTIL ([Bmpy][NTf2]) in the solid polymer electrolyte after high-temperature
casting was not denatured according to FTIR of Figure c. It was found that the solid polymer electrolyte
and RTIL are consistent with the characteristic peaks. According to
the absorption peak of PVDF and the characteristic band of the α
phase, it indicated that PVDF appeared as the α phase.
Figure 3
(a) XRD and
(b) XPS analysis of Pt-loaded Ti foam, (c) FTIR of
RTIL and solid polymer electrolyte, and (d) determination of hydrogen
adsorption and desorption peaks in CV of Pt-loaded Ti foam in 0.1
mmol/L perchloric acid solution.
(a) XRD and
(b) XPS analysis of Pt-loaded Ti foam, (c) FTIR of
RTIL and solid polymer electrolyte, and (d) determination of hydrogen
adsorption and desorption peaks in CV of Pt-loaded Ti foam in 0.1
mmol/L perchloric acid solution.In our previous work, the large porosity of the porous ceramic
was used as the substrate and the Pt SPE has a rough surface,[21,24] which results in a small three-phase interfacial area between the
smooth solid polymer electrolyte membrane and WE. The Pt electrode
layer with a thickness of 10 μm was formed, which led to a waste
of precious metal. Bulk Pt electrodes not only increase the fabrication
cost but also reduce the sensitivity of the sensor due to the small
electrochemically active surface area (ECSA) and low catalytic efficiency.
Conversely, the three-dimensional structure of the metal foam provides
a high specific surface area for the loading of Pt-NPs. The ECSA of
the material was characterized by CV in 0.1 mmol/L perchloric acid
solution.[25,26] It found that the adsorption and desorption
peaks of hydrogen appeared around −0.1 V, as shown in Figure d. The charges for
the hydrogen adsorption and desorption processes were obtained by
integration to be 6.85 mC and 5.75 mC. According to the formulaQ is the electric quantity
(unit, C), and C is the unit adsorption capacitance
of Pt to hydrogen, which was a constant of 0.21 mC/cm2.
The resulting ECSA was 32.6 cm2. It was shown that the
Pt-supported Ti foam electrode had great electrochemical activity.
Sensing Properties of Hydrogen
[Bmpy][NTf2] is more suitable as the electrolyte for HOR by comparing
different RTILs. According to Silvester et al., RTIL based on [NTf2]− can form a stable protonated product
[HNTF2] in the HOR with a good electrochemical stability
and a wide electrochemical potential window.[27,28] Meanwhile, compared with [Bmim][NTf2], [Bmpy][NTf2] has lower oxidation peak voltage, higher oxidation peak
current, and higher hydrogen solubility, indicating that [Bmpy][NTf2] has a more reasonable RTIL-electrode interface structure.
CV needs to limit the gas diffusion velocity as much as possible to
achieve diffusion-controlled conditions. At this time, the electrode
reaction mainly depends on the HOR, which is beneficial to better
explore the reaction mechanism. Due to the high viscosity of RTIL,
it can be used as a diffusion barrier. Hydrogen dissolved and diffused
in RTIL, followed by hydrogen adsorption and oxidation reactions at
the three-phase interface. Finally, the hydrophobic polytetrafluoroethylene
(PTFE) film was designed to block water vapor, which could cause cross-sensitive
response to sensors. In addition, it could also allow air to flow
smoothly when exposed to analyte gases. Due to the wide application
of LIBs in life, transportation, aerospace, and other fields, it is
necessary to detect hydrogen under aerobic and anaerobic conditions
to ensure the safety of LIBs.
Sensing Properties under
Anaerobic Conditions
CV was used to study the HOR mechanism
under anaerobic conditions,
and the scanning voltage range was set to −0.6–1.0 V.
It is important to select a suitable scan rate that matches the gas
flow rate for CV. When the scan rate decreases, the extent of the
diffusion layer increases, the polarization current decreases accordingly,
and the peak value of the CV curve becomes more prominent and obvious.
However, the electrode reaction rate also decreases accordingly, which
may cause the reaction to enter the reaction-controlled zone. Therefore,
the selected scan rate was 20 mV/s during the CV. When the hydrogen
concentration increases in turn, after completing the CVs under different
hydrogen concentrations, the experimental data are shown in Figure a. The oxidation
peak voltage was 0.35 V, reduction peak voltage was −0.4 V,
and E1/2 was −0.025 V, which was
quite close to zero potential, as shown in Figure a.
Figure 4
(a) CV of SPE at different hydrogen concentrations
under anaerobic
conditions, (b) chronoamperometry of the “sandwich”
sensor at different hydrogen concentrations, and the insets are the
linear fit of current and hydrogen concentration.
(a) CV of SPE at different hydrogen concentrations
under anaerobic
conditions, (b) chronoamperometry of the “sandwich”
sensor at different hydrogen concentrations, and the insets are the
linear fit of current and hydrogen concentration.It was also found that the oxidation peak voltage shifted to the
right and the reduction peak voltage shifted to the left with the
increase of hydrogen concentration, while the E1/2 remained basically unchanged. When hydrogen concentration
increased, a larger forward voltage was required correspondingly to
ensure that the reaction rate was fast enough to deplete the hydrogen.
The concentration of hydrogen and proton on the electrode surface
was the same when E = E1/2,[29] and the oxidation and reduction reaction
rates reached a temporary balance. Then, the outward diffusion rate
of proton was accelerated. It can be observed that when the hydrogen
concentration was greater than 6000 ppm, the oxidation peak potential
shifted to the right, and the reason for the left shift of the reduction
peak was basically the same as that of the oxidation peak. In addition,
the reduction peak current of the CV curve was found to be larger
than the oxidation peak. WE is a Pt-loaded Ti foam and CE is a carbon
electrode. Thus, the electrochemical activity of CE is much smaller
than that of WE. When the CV was scanning, the generated protons cannot
react completely and gradually accumulated. Furthermore, the accumulated
protons in the negative CV scan all react on WE, resulting in the
reduction reaction kinetics which was greater than that of the oxidation
reaction. With the progress of each CV, the reduction current was
increasing, leading to the unclosed curve. Within the range of 10 000
ppm, it indicated that the electrochemical reaction was under diffusion-controlled
conditions. The linear fitting between the hydrogen concentration
and the peak currents of oxidation and reduction reached 0.99 and
0.98, respectively, showing a good linear relationship, as shown in Figure a. However, considering
the phenomenon of the right shift of the peak potential, the hydrogen
concentration was adjusted to be less than 6000 ppm in the chronoamperometry
process, which was more reliable.
Figure 5
(a) CV for different hydrogen concentrations
under aerobic conditions
and (b) chronoamperometry at different hydrogen concentrations. The
insets illustrate the linear fit of the chronoamperometry current
to the hydrogen concentration.
(a) CV for different hydrogen concentrations
under aerobic conditions
and (b) chronoamperometry at different hydrogen concentrations. The
insets illustrate the linear fit of the chronoamperometry current
to the hydrogen concentration.It is well known that HOR in acidic aqueous electrolytes basically
follows the Tafel–Volmer and Heyrovsky–Volmer pathways
processes.[30] The steps of HOR of Ti-based
Pt electrodes in RTIL are similar to the Tafel reaction, Volmer reaction,
and Heyrovsky reaction.[31] Hydrogen was
firstly adsorbed on the Pt surface to form Pt–H(ad), and Pt–H(ad) was oxidized to form a hydrogen
radical (Pt–H•) and a proton (H+) in the second step, which followed eqs –4When performing
chronoamperometry experiments, the performance
and lifetime of the electrochemical sensor have a great dependence
on the setting of potential. According to the oxidation peak potential
obtained by CV, 0.5 V was selected as the bias voltage. The gas diffused
simultaneously on both sides through Ti foam and porous ceramic and
reached the three-phase interface directly, without the diffusion
process in RTIL. The sensor package was designed with a vent, and
a through gas diffusion path was formed during the sensor detection
process, as shown in the layered device structure of Figure c. The sensitivity of the sensor
reached 3.9 nA/ppm, and the T90 of the
sensor was about 92 s (Figure b). The noise of the sensor was 4 nA, and the detection limit
(LOD) was calculated to be 3.1 ppm based on a signal-to-noise ratio
(SNR) of 3. In addition, the linear fitting value of hydrogen concentration
and response current was 0.99, which had a good linear relationship
(Figure b).In fact, the reduction reaction rate on CE is independent of the
oxidation reaction rate on WE. Because it was found that there is
an open circuit between WE and CE, no charge transfer occurs. CE only
acts to balance the voltage drop produced by the Faradaic current
and reduce the oxidation products, so as to avoid excessive accumulation
of the proton in RTIL in the HOR. In turn, the electrolyte will be
damaged and the lifespan of the sensor will be shortened by excessive
acidification for a long time. The protons released by WE need to
diffuse in RTIL and migrate to CE for reduction reaction. However,
WE and CE form a conducting loop in the traditional three-electrode
principle of electrochemistry, and the reaction rates on WE and CE
are consistent. Therefore, different from the traditional three-electrode
principle in electrochemistry, the migration rate and reduction reaction
rate of protons in RTIL of sensors under applied bias voltage do not
limit the electrochemical reaction rate on WE.
Sensing Properties under Aerobic Conditions
Next, the
mechanism of HOR under aerobic conditions was explored
using CV. The scanning voltage range is −1.5 to 1.8 V, and
the potential range is wider than that under anaerobic conditions.
Therefore, the drying of RTIL and the removal of gaseous and water
molecules are important, otherwise additional peaks may appear in
the water electrolysis reaction at high potential. In the negative
voltage sweep, oxygen is reduced and superoxide radical (O2•–) is generated, unlike in anaerobic conditions,
as shown in eq . During the positive voltage sweep, O2•– is oxidized and Pt–O is generated, which is seen in eq . The CV curve shows
a reduction peak at −0.75 V and an oxidation peak at 0.3 V,
which is shown in the results of CV in synthetic air in Figure a.In addition, it could
be observed that oxygen is consumed by HOR under aerobic conditions.
The reason is that the current decreases at the oxidation peak at
0.3 V when exposed to hydrogen. At the same time, with the increase
in hydrogen concentrations, the reduction current increases slightly
around −1 V, suggesting that there is a coupled reaction between
the hydrogen oxidation process and the oxygen reduction process. The
higher the hydrogen concentration, the more O2•– that is consumed by HOR. Consequently, the oxygen reduction process
is promoted according to eq . It can be seen that the mechanism of HOR is different
from that under anaerobic conditions due to the participation of oxygen.
The oxidation peak of HOR appears around 1.2 V, which is higher than
the oxidation peak potential under anaerobic conditions. It shows
that there is a good linear relationship between the peak current
and gas concentration in the range of 5000 ppm. The linear fitting
values between hydrogen concentration and the peak currents of oxidation
and reduction of CV reached 0.99 and 0.98, respectively, as shown
in Figure a. It is
indicated that the electrochemical reaction is in the diffusion-controlled
region in the hydrogen concentration range of 0–5000 ppm. The
reaction mechanism is shown below, which was first proposed by Zeng[31]The chronoamperometry
experiment was performed in aerobic conditions
with a bias voltage of 1 V which was based on the oxidation peak potential
of the HOR by CV (Figure a). It was found that the current of the HOR under aerobic
conditions was smaller and the sensitivity was lower than that in
the anaerobic environment. The sensitivity of the sensor was 0.9 nA/ppm,
the T90 was 84 s, the noise was 3 nA,
and LOD was 6.6 ppm, as shown in the results of Figure b. The electrochemical reaction under diffusion-controlled
conditions was also illustrated in the hydrogen concentration range
of 5000 ppm. The linear fit of the current under different oxygen
concentrations reached about 0.99. At the same time, it was found
that there was another mechanism of HOR in addition to eqs –9 in aerobic conditions. It can be observed that
the baseline current continuously moved down with the increase of
hydrogen concentration, as shown in Figure b. The product (Pt–HO2•) of the coupling reaction of the hydrogen oxidation
process and the oxygen reduction process could continue to participate
in the oxidation or reduction reaction.[31]Equation indicated
that the protons were also generated at the same time as the hydrogen
radical intermediate (Pt–H•). The kinetics
of the reduction reaction of Pt–HO2• was greatly enhanced due to the accumulation of protons after the
hydrogen response phase, as shown in eq , which led to a partial cancellation of
the positive background current in the equilibrium phase. It is even
foreseeable that if the hydrogen concentration is increased, the background
current may become negative. Therefore, electrochemical gas sensors
are not suitable for detecting high concentrations of hydrogen in
aerobic conditions and should be avoided, as reported for common electrochemical
gas sensors.[31]Furthermore, the tolerance
of temperature on the sensor performance
was investigated. A series of temperatures were controlled within
the range of 243–323 K, and the output current of the sensor
was detected under the same hydrogen concentration (2000 ppm). When
the temperature was lower than 243 K, the sensor lost response due
to the limited mass transfer process and the extremely slow reaction
rate. At lower temperatures (243–273 K), both the response
speed and the sensitivity of the sensor become smaller, the background
current is larger, as shown in Figure a. It showed that the hydrogen dissolved in RTIL was
not completely reacted due to the slow rate of HOR. When the temperature
between 273 and 323 K, the background current is significantly reduced
and better sensitivity is achieved. Above 323 K, the electrical connection
degraded and the sensor output became unstable. The accumulation of
excess reaction by-products in the electrolyte film can lead to changes
in the electrical properties, such as conductivity or acidity. As
a result, the sensitivity of the sensor changes and the signal of
sensor becomes unreliable. In severe cases, it will directly deplete
the electrolyte and reduce the life of the sensor.[32] The relationship between the temperature of the electrochemical
sensor and the output current conforms to the Arrhenius formula,[33] as shown in eq Here, k is the rate constant, R is the molar gas constant, T is the thermodynamic
temperature, Ea is the apparent activation
energy, and B is the frequency factor. It is also
indicated that the sensor can work normally in the temperature range
of 243–323 K. The response at each temperature was fitted,
and the linearity was found to be 0.99, as shown in Figure a. In addition, it shows that
the stability and repeatability of the sensor meet the standards.
The transient response to hydrogen (300 ppm balanced with N2) was repeated for 8 h as given in Figure b. It was found that the response of the
sensor decreased a bit during the test. The reason is speculated to
be the slight accumulation of protons during long-term testing according
to the electrochemical behavior of hydrogen mentioned above, which
in turn leads to a slight decrease in the HOR kinetics of WE. Overall,
one could find that both the base and responding currents have a very
small drift less than 0.3 nA for a long time, which indicates our
sensors have a very good stability and reproducibility.
Figure 6
(a) Sensing
performance under different temperature conditions,
with the linear fit of the response current in the inset, (b) prolonged
continuous operation (48 cycles within 8 h), and (c) magnified images
of three of the response stages in part b.
(a) Sensing
performance under different temperature conditions,
with the linear fit of the response current in the inset, (b) prolonged
continuous operation (48 cycles within 8 h), and (c) magnified images
of three of the response stages in part b.Finally, the sandwich device structure scheme provides a good idea
for the development of high-performance electrochemical gas sensors
and is also applicable to other porous electrode materials. Two different
materials, nickel foam (Ni foam) and copper foam (Cu foam), were used
to load Pt. The characterization of the materials and the exploration
of the electrochemical behavior of hydrogen are shown in Figure S4. The feasibility of different electrode
materials for electrochemical hydrogen sensors is preliminarily shown.
It also reflects the potential research value of the sandwich device
structure.
Conclusion
In summary, we proposed
an electrochemical gas sensor design with
a “sandwich” structure of Pt-NPs-supported metal foam
and SPE electrodes based on porous substrates. The solid polymer electrolyte
was innovatively filled into the porous WE, which optimizes the gas
diffusion path and effectively improves the sensitivity and response
speed of the sensor. The HOR mechanism of the Pt-loaded Ti foam electrode
under both anaerobic and aerobic conditions was fully discussed, and
the high performance of the sensor was demonstrated. The development
of the sensor provides a commercial idea for the safe real-time monitoring
of LIBs. In the next step,
the potentiostatic hydrogen detector will conduct larger-scale experiments
in a practical application environment and extend the “sandwich”
device structure to a wider range of detection of different kinds
of gases.
Authors: Zhenguo Yang; Jianlu Zhang; Michael C W Kintner-Meyer; Xiaochuan Lu; Daiwon Choi; John P Lemmon; Jun Liu Journal: Chem Rev Date: 2011-03-04 Impact factor: 60.622