| Literature DB >> 8413008 |
Abstract
The image information transfer properties of a number of x-ray fluorescent screens have been measured for x-ray energies from 17 to 320 keV. The detective quantum efficiency of the screens at each x-ray energy has been determined by separate measurements of the x-ray absorption efficiency and the statistical factor associated with the emission of optical photons upon absorption of an incident x-ray. Data have been recorded for both rare-earth phosphor screens and calcium tungstate screens. The value of the statistical factor for optical photon emission tends toward a constant value as the incident energy increases. Comparisons of the image information transfer properties are presented for several screens, which have been measured over a ten year interval. The utility of the screens for high-energy radiography is discussed.Entities:
Mesh:
Year: 1993 PMID: 8413008 DOI: 10.1118/1.596998
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071