Evaluation of a transparent blood analog fluid: aqueous xanthan gum/glycerin.

K A Brookshier1, J M Tarbell.   

Abstract

Aqueous solutions of Xanthan gum and glycerin were developed that match both the viscous and elastic components of the 2 Hz complex viscosity of blood over a broad range of shear rates (1-1000 s-1) and hematocrits (20%-80%). Measurements of wall shear rate in pulsatile flow were conducted in straight and atherogenic curved artery models using both the blood analog fluid and blood to compare their flow behavior. Peak wall shear rates determined in these models under nearly identical flow conditions with the two fluids were not statistically different (p > 0.05). These results indicate that aqueous Xanthan gum/glycerin solutions provide a good blood analog fluid for flow conditions characteristic of large arteries. Because the fluid is transparent and made from readily available and fairly inexpensive materials, it may be generally useful for in vitro hemodynamic studies, particularly those employing optical techniques.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8400149     DOI: 10.3233/bir-1993-30202

Source DB:  PubMed          Journal:  Biorheology        ISSN: 0006-355X            Impact factor:   1.875


  7 in total

1.  Extensional flow of blood analog solutions in microfluidic devices.

Authors:  P C Sousa; F T Pinho; M S N Oliveira; M A Alves
Journal:  Biomicrofluidics       Date:  2011-03-17       Impact factor: 2.800

2.  Flow-induced deformation in a microchannel with a non-Newtonian fluid.

Authors:  Kiran Raj M; Jeevanjyoti Chakraborty; Sunando DasGupta; Suman Chakraborty
Journal:  Biomicrofluidics       Date:  2018-06-25       Impact factor: 2.800

3.  The influence of operational protocol on the fluid dynamics in the 12 cc Penn state pulsatile pediatric ventricular assist device: the effect of end-diastolic delay.

Authors:  Benjamin T Cooper; Breigh N Roszelle; Tobias C Long; Steven Deutsch; Keefe B Manning
Journal:  Artif Organs       Date:  2010-04       Impact factor: 3.094

4.  Hemodynamic characteristics in a cerebral aneurysm model using non-Newtonian blood analogues.

Authors:  Hang Yi; Zifeng Yang; Mark Johnson; Luke Bramlage; Bryan Ludwig
Journal:  Phys Fluids (1994)       Date:  2022-10-03       Impact factor: 4.980

5.  Simulating Radial Pressure Waveforms with a Mock Circulatory Flow Loop to Characterize Hemodynamic Monitoring Systems.

Authors:  Anna Packy; Gavin A D'Souza; Masoud Farahmand; Luke Herbertson; Christopher G Scully
Journal:  Cardiovasc Eng Technol       Date:  2021-09-01       Impact factor: 2.305

6.  Hemodynamic Performance of a Novel Right Ventricular Assist Device (PERKAT).

Authors:  Daniel Kretzschmar; P Christian Schulze; Markus W Ferrari
Journal:  ASAIO J       Date:  2017 Mar/Apr       Impact factor: 2.872

7.  Hemodynamic diagnostics of epicardial coronary stenoses: in-vitro experimental and computational study.

Authors:  Rupak K Banerjee; Koustubh D Ashtekar; Tarek A Helmy; Mohamed A Effat; Lloyd H Back; Saeb F Khoury
Journal:  Biomed Eng Online       Date:  2008-08-27       Impact factor: 2.819

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.