Simulating Radial Pressure Waveforms with a Mock Circulatory Flow Loop to Characterize Hemodynamic Monitoring Systems.

Anna Packy1,2, Gavin A D'Souza1, Masoud Farahmand1, Luke Herbertson1, Christopher G Scully3.   

Abstract

PURPOSE: Mock circulatory loops (MCLs) can reproducibly generate physiologically relevant pressures and flows for cardiovascular device testing. These systems have been extensively used to characterize the performance of therapeutic cardiac devices, but historically MCLs have had limited use for assessing patient monitoring systems. Here, we adapted an MCL to include peripheral components and evaluated its utility for qualitative and quantitative benchtop testing of hemodynamic monitoring devices.
METHODS: An MCL was designed to simulate three physiological hemodynamic states: normovolemia, cardiogenic shock, and hyperdynamic circulation. The system was assessed for stability in pressure and flow values over time, repeatability, waveform morphology, and systemic-peripheral pressure relationships.
RESULTS: For each condition, cardiac output was controlled to the nearest 0.2 L/min, and flow rate and mean arterial pressure remained stable and repeatable over a 60-s period (n = 5, standard deviation of ± 0.1 L/min and ± 0.84 mmHg, respectively). Transfer function analyses showed that the systemic-peripheral relationships could be adequately manipulated. The results from this MCL were comparable to those from other published MCLs and computational simulations. However, resolving current limitations of the system would further improve its utility. Three pulse contour analysis algorithms were applied to the pressure and flow data from the MCL to demonstrate the potential role of MCLs in characterizing hemodynamic monitoring systems.
CONCLUSION: Overall, the development of robust analysis methods in conjunction with modified MCLs can expand device testing applications to hemodynamic monitoring systems. Properly validated MCLs can create a stable and reproducible environment for testing patient monitoring systems over their entire operating ranges prior to clinical use.
© 2021. This is a U.S. government work and not under copyright protection in the U.S.; foreign copyright protection may apply.

Entities:  

Keywords:  Cardiac device testing; Cardiovascular flow circuit; Hemodynamic monitoring devices; Pulse contour analysis

Mesh:

Year:  2021        PMID: 34472042      PMCID: PMC8885788          DOI: 10.1007/s13239-021-00575-2

Source DB:  PubMed          Journal:  Cardiovasc Eng Technol        ISSN: 1869-408X            Impact factor:   2.305


  42 in total

Review 1.  Validation of the transfer function technique for generating central from peripheral upper limb pressure waveform.

Authors:  David Gallagher; Audrey Adji; Michael F O'Rourke
Journal:  Am J Hypertens       Date:  2004-11       Impact factor: 2.689

2.  A novel study of mechanical heart valve cavitation in a pressurized pulsatile duplicator.

Authors:  Changfu Wu; Stephen M Retta; Ronald A Robinson; Bruce A Herman; Laurence W Grossman
Journal:  ASAIO J       Date:  2009 Sep-Oct       Impact factor: 2.872

3.  Evaluation of radial and ulnar blood flow after radial artery cannulation with 20- and 22-gauge cannulae using duplex Doppler ultrasound.

Authors:  S Y Kim; J S Lee; W O Kim; J M Sun; M K Kwon; H K Kil
Journal:  Anaesthesia       Date:  2012-07-16       Impact factor: 6.955

4.  Development of hyperdynamic circulation and response to β-blockers in compensated cirrhosis with portal hypertension.

Authors:  Càndid Villanueva; Agustín Albillos; Joan Genescà; Juan G Abraldes; Jose L Calleja; Carles Aracil; Rafael Bañares; Rosa Morillas; María Poca; Beatriz Peñas; Salvador Augustin; Joan Carles Garcia-Pagan; Oana Pavel; Jaume Bosch
Journal:  Hepatology       Date:  2015-11-26       Impact factor: 17.425

Review 5.  Accuracy of stroke volume variation in predicting fluid responsiveness: a systematic review and meta-analysis.

Authors:  Zhongheng Zhang; Baolong Lu; Xiaoyan Sheng; Ni Jin
Journal:  J Anesth       Date:  2011-09-04       Impact factor: 2.078

6.  Development and Validation of a Life-Sized Mock Circulatory Loop of the Human Circulation for Fluid-Mechanical Studies.

Authors:  Johannes Gehron; Julian Zirbes; Markus Bongert; Stefan Schäfer; Martin Fiebich; Gabriele Krombach; Andreas Böning; Philippe Grieshaber
Journal:  ASAIO J       Date:  2019 Nov/Dec       Impact factor: 2.872

7.  Evaluation of a transparent blood analog fluid: aqueous xanthan gum/glycerin.

Authors:  K A Brookshier; J M Tarbell
Journal:  Biorheology       Date:  1993 Mar-Apr       Impact factor: 1.875

Review 8.  An analysis of the relationship between central aortic and peripheral upper limb pressure waves in man.

Authors:  M Karamanoglu; M F O'Rourke; A P Avolio; R P Kelly
Journal:  Eur Heart J       Date:  1993-02       Impact factor: 29.983

9.  Evaluation of pulmonary artery stenosis in congenital heart disease patients using functional diagnostic parameters: An in vitro study.

Authors:  Gavin A D'Souza; Rupak K Banerjee; Michael D Taylor
Journal:  J Biomech       Date:  2018-09-15       Impact factor: 2.712

10.  In Vitro Validation of a Multiscale Patient-Specific Norwood Palliation Model.

Authors:  Tianqi Hang; Alessandro Giardini; Giovanni Biglino; Timothy Conover; Richard S Figliola
Journal:  ASAIO J       Date:  2016 May-Jun       Impact factor: 2.872

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.