Comparative accuracy of six ambulatory devices according to blood pressure levels.

E O'Brien1, N Atkins, F Mee, K O'Malley.   

Abstract

OBJECTIVE: To assess the accuracy of six ambulatory blood pressure measuring systems at low, medium and high blood pressures.
RESULTS: The CH-Druck, Profilomat, SpaceLabs 90207 and Novacor DIASYS 200R, having previously achieved A to C grading for systolic and diastolic blood pressures according to the British Hypertension Society (BHS) protocol and having fulfilled the criteria of the Association for the Advancement of Medical Instrumentation, have been recommended for measurement of ambulatory blood pressure in clinical practice; the Pressurometer IV and Takeda TM-2420, achieved only C and D grades and failed to satisfy the Association for the Advancement of Medical Instrumentation criteria. In this study the data from the original validations are re-analysed for three pressure ranges of systolic and diastolic blood pressures: low range < or = 130/80 mmHg, medium range 130-160/80-100 mmHg and high range > or = 160/100 mmHg. All six devices maintained their overall grading or improved them slightly in the low and medium blood pressure ranges, but in the high blood pressure range the CH-Druck slipped from an overall A/A grading to B/C, the Profilomat from B/A to C/D, the SpaceLabs from B/B to C/C and the Pressurometer IV from C/D to D/D. The Takeda remained unchanged with a D grading, but the results within this grading were worse in the higher blood pressure range, and the Novacor rose from C/C to C/B.
CONCLUSIONS: This analysis suggests that the CH-Druck is the most accurate ambulatory system across the pressure range, although it does not perform as well in the high blood pressure range as in the medium and low blood pressure ranges. The SpaceLabs 90207 is accurate in the low and medium blood pressure ranges and reasonably accurate in the high blood pressure range. If blood pressures only in the low and medium ranges are to be measured, a wider selection of ambulatory systems becomes available because, in addition to the CH-Druck and SpaceLabs 90207, the Profiloat and Novacor DIASYS 200R are accurate.

Entities:  

Mesh:

Year:  1993        PMID: 8397247     DOI: 10.1097/00004872-199306000-00012

Source DB:  PubMed          Journal:  J Hypertens        ISSN: 0263-6352            Impact factor:   4.844


  7 in total

1.  Nocturnal systemic hypotension increases the risk of glaucoma progression.

Authors:  Mary E Charlson; Carlos Gustavo de Moraes; Alissa Link; Martin T Wells; Gregory Harmon; Janey C Peterson; Robert Ritch; Jeffrey M Liebmann
Journal:  Ophthalmology       Date:  2014-05-25       Impact factor: 12.079

2.  Relation between insufficient response to antihypertensive treatment and poor compliance with treatment: a prospective case-control study.

Authors:  R Nuesch; K Schroeder; T Dieterle; B Martina; E Battegay
Journal:  BMJ       Date:  2001-07-21

3.  Agreement between the Takeda UA-731 automatic blood pressure measuring device and the manual mercury sphygmomanometer: an assessment under field conditions in Newcastle upon Tyne, UK.

Authors:  C Cartwright; N Unwin; P Stephenson
Journal:  J Epidemiol Community Health       Date:  1996-04       Impact factor: 3.710

4.  24 hour blood pressure control with once-daily versus twice-daily formulations of diltiazem.

Authors:  T D Ruddy; J M Wright; D Savard; S P Handa; A Chockalingam; L Fischer; A P Boulet
Journal:  Cardiovasc Drugs Ther       Date:  1995-12       Impact factor: 3.727

5.  Limitations of the of the oscillometric method for blood pressure measurements in dialyzed patients.

Authors:  Marek Czarkowski; Monika Staszkow; Kacper Kostyra; Zuhier Shebani; Krzysztof Rozanowski; Joanna Matuszkiewicz-Rowinska; Stanislaw Niemczyk
Journal:  Med Sci Monit       Date:  2011-04

6.  The impact of central blood pressure levels on the relationship between oscillometric and central blood pressure measurements: a multicenter invasive study.

Authors:  Yusuf I Alihanoglu; Mehmet Kayrak; Mehmet S Ulgen; Mehmet Yazici; Mehmet Yazici; Remzi Yilmaz; Kenan Demir; Yildiz Dogan; Murat Sizer; Hakan Ozhan; Fatih Koc; Sait Bodur
Journal:  J Clin Hypertens (Greenwich)       Date:  2013-07-16       Impact factor: 3.738

7.  Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension.

Authors:  Jalil Belghazi; Ramzi N El Feghali; Thérèse Moussalem; Maya Rejdych; Roland G Asmar
Journal:  Vasc Health Risk Manag       Date:  2007
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.