Characteristics of intracardiac electrograms.

G H Myers, Y M Kresh, V Parsonnet.   

Abstract

Cardiac electrograms were measured on endocardial leads on patients undergoing pacemaker implantation. The data show that small area electrodes reduce R wave amplitude, decrease width of the complex, and increase slopes. The bandwidth of the electrogram increases as the electrode area decreases. The loading of the sensing amplifier changes these characteristics. Distortion is most pronounced for small area electrodes and low lead resistances.

Entities:  

Mesh:

Year:  1978        PMID: 83626     DOI: 10.1111/j.1540-8159.1978.tb03445.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  Polarisation impedance of pacemaker electrodes: in vitro studies simulating practical operation.

Authors:  H Fischler; H P Schwan
Journal:  Med Biol Eng Comput       Date:  1981-09       Impact factor: 2.602

2.  T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing.

Authors:  Ya-Xun Sun; Jing Gao; Chen-Yang Jiang; Yu-Mei Xue; Yi-Zhou Xu; Gang Liu; Ji-Hong Guo; Xia Sheng; Yang Ye; Hong He; Yun-Tao Zhao; Hector Barajas-Martinez; Guo-Sheng Fu; Dan Hu
Journal:  Front Physiol       Date:  2017-09-01       Impact factor: 4.566

3.  What is the mechanism of irregular pacing spikes? Pacing failure, sensing failure, or something else? A case report.

Authors:  Gaku Oguri; Toshiya Kojima; Katsuhito Fujiu; Issei Komuro
Journal:  Eur Heart J Case Rep       Date:  2020-04-04

4.  Time course of current of injury is related to acute stability of active-fixation pacing leads in rabbits.

Authors:  Shalaimaiti Shali; Alimujiang Wushou; Entao Liu; Lin Jia; Ruiming Yao; Yangang Su; Junbo Ge
Journal:  PLoS One       Date:  2013-03-05       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.