| Literature DB >> 28919866 |
Ya-Xun Sun1, Jing Gao1, Chen-Yang Jiang1, Yu-Mei Xue2, Yi-Zhou Xu3, Gang Liu4, Ji-Hong Guo5, Xia Sheng1, Yang Ye1, Hong He1, Yun-Tao Zhao6, Hector Barajas-Martinez7, Guo-Sheng Fu1, Dan Hu7,8.
Abstract
Introduction: T wave oversensing (TWOS) is a major drawback of implantable cardioverter defibrillator (ICD) and data on predictors of TWOS in ICD is limited. We aimed to calculate a novel index of T wave safety margin (TWSM) and assess its potential for evaluating TWOS during the procedure of ICD implantation. Methods andEntities:
Keywords: T wave oversensing; implantable cardioverter-defibrillator; inappropriate therapy; intracardiac electrocardiogram; safety margin
Year: 2017 PMID: 28919866 PMCID: PMC5585188 DOI: 10.3389/fphys.2017.00659
Source DB: PubMed Journal: Front Physiol ISSN: 1664-042X Impact factor: 4.566
Demographic and clinical characteristics of patients.
| Age, y/o (Mean ± | 54.9 ± 12.3 |
| Male, | 19 (59%) |
| Primary prevention, | 23 (72%) |
| Secondary prevention, | 9 (28%) |
| Dilated cardiomyopathy | 10 (31%) |
| Reduced LVEF post-myocardial infarction | 6 (19%) |
| Survived idiopathic cardiac arrest | 5 (16%) |
| Brugada syndrome | 4 (13%) |
| Hypertrophic cardiomyopathy | 3 (9%) |
| Long-QT syndrome | 2 (6%) |
| Alcoholic cardiomyopathy | 1 (3%) |
| Short-QT syndrome | 1 (3%) |
| LVEF, % ( | 48 ± 19 (32) |
| Dilated cardiomyopathy | 31 ± 3(10) |
| Ischemic cardiomyopathy | 33 ± 2 (6) |
| Hypertrophic cardiomyopathy | 76 ± 5 (3) |
| Arrhythmias related | 62 ± 10 (13) |
| Serum potassium (mmol/L) | 3.9 ± 0.3 |
ICD, Implantable cardioverter-defibrillator; LVEF, left ventricular ejection fraction.
Figure 1(A) Ex vivo ICD connection during the procedure of ICD implantation using the active-fixation lead. (B) TWSM measurements. The green line indicates the sensing threshold curve of ICD. (C) The timeline for the evaluation of TWSM.
Figure 2Bland-Altman analysis comparing R wave amplitudes measured from implantable cardioverter-defibrillator printouts. (A) Measurement of R wave amplitude by Fortify ICD. Noted that IEGM of ICD automatically switched to “Ventricular amplifier” channel to measure R wave amplitude. (B,C) R wave amplitude measured from by Fortify ICD. (D,E) R wave amplitude measured by Virtuoso ICD. Inner dashed line is the mean difference; outer dashed lines are mean ± 1.96 SD.
Figure 3Comparison of R wave amplitude, T wave amplitudes and TWSMs by different ICDs. (A). R wave amplitudes by different ICDs; (B,C) T wave amplitudes post sensing intrinsic rhythm and post pacing at 90/min in Fortify and Teligen ICDs. (D) TWSM after sensing intrinsic rhythm and pacing in Fortify and Teligen. Red line indicates safety margin of 2 (200%). *P < 0.05; **P < 0.01.
Details of patients with TWOS during the procedure of ICD implantation.
| 1 | 54, M | SQTS | Primary | + | − | − | − | − | − | Vitality | Riata 1570 |
| 2 | 56, M | Brugada | Secondary | − | + | ± | − | − | − | Fortify | Durata 7122 |
| 3 | 42, F | HCM | Primary | − | + | − | − | − | − | Fortify | Durata 7122 |
| 4 | 65, M | DCM | Primary | − | ± | ± | − | − | − | Current | Durata 7122 |
| 5 | 28, M | Brugada | Secondary | − | + | − | − | − | ± | Teligen | RELIANCE G |
| 6 | 52, F | SCD | Secondary | − | + | + | − | − | − | Fortify | Durata 7122 |
| 7 | 44, F | HCM | Primary | + | − | ± | − | − | − | Teligen | RELIANCE G |
| 8 | 58, M | Brugada | Primary | − | + | − | − | − | − | Current | Durata 7122 |
| 9 | 54, M | Post MI | Primary | − | + | + | − | + | − | Fortify | Durata 7122 |
| 10 | 47, F | DCM | Primary | − | + (BV) | − | − | − | − | Consulta | Sprint 6947 |
BV, Bioventricular pacing; ICD, implantable cardioverter-defibrillator; TWOS, T-wave oversensing; VP, ventricular pacing; VS, ventricular sensing; PSVP, post-sensing ventricular pacing. +, persistent TWOS; −, no TWOS; ±, intermittent TWOS;
during routine follow-up.
Figure 4TWOS episodes during the procedure of ICD implantation and sensing threshold curves with real proportion. (A) 12-lead surface electrocardiogram of the short QT syndrome patient. (B) TWOS under the maximum sensitivity in Virtuoso ICD. Red line shows the blank period after sensing; green line shows the sensing curve with minimum sensitivity; blue lines show sensing curves with sensitivity of 0.3 and 0.15 mV; red dot shows the possible range of T wave amplitudes. (C) Intermittent TWOS under maximum sensitivity in Epic ICD. Green and blue lines show the sensing curves with minimum and default sensitivity, respectively. (D) Measurement of TWSM in Fortify ICD. (E,F) Measurement of TWSM in Vitality and Teligen ICDs. Blue lines show the sensing curve with default sensitivity.
Figure 5Post pacing TWOS revealed during the procedure of ICD implantation in Brugada syndrome patient, hypertrophic cardiomyopathy patient and dialated cardiomyopathy patient. (A) 12-lead surface EKG of a Brugada syndrome patient. (B,C) Post pacing TWOS episodes by a Consulta ICD under sensitivity of 0.15 mV. No TWOS was recorded under 0.3 mV. (D) 12-lead surface EKG of a hypertrophic cardiomyopathy patient. The implantable cardioverter-defibrillator working mode is VAT. (E) Post pacing TWOS episodes in Virtuoso ICD under the sensitivity of 0.15 mV. (F) 12-lead EKG of a CRT-D patient with dilated cardiomyopathy. (G) After the episode of TWOS (①), PVARP was prolonged by PVC response and sinus P wave was recognized as a premature beat (②) and not traced by ventricle (③).