| Literature DB >> 8353942 |
A Andersson1, A Isaksson, L Brattström, B Hultberg.
Abstract
We describe a versatile high-performance liquid-chromatographic method for determining homocysteine and other plasma sulfhydryls. Using three different procedures for preparation of plasma, we determined total, free (non-protein-bound), and reduced forms of homocysteine, cysteine, glutathione, cysteinylglycine, and gamma-glutamylcysteine in human plasma. Sample preparation involves disulfide reduction with dithiothreitol and protein precipitation with sulfosalicylic acid. The assay utilizes isocratic reversed-phase ion-pair liquid chromatography at pH 2.4, postcolumn derivatization with 4,4'-dithiodipyridine, and colorimetric detection at 324 nm. The intra-assay precision (CV) of the method for total homocysteine is 1.5%; the interassay precision over 2.5 months is 2.5%. The detection limit for homocysteine is < 50 nmol/L plasma.Entities:
Mesh:
Substances:
Year: 1993 PMID: 8353942