Literature DB >> 8347025

Intraaortic balloon entrapment.

M D Horowitz1, M Otero, E J de Marchena, R M Neibart, S Novak, H Bolooki.   

Abstract

A review of intraaortic balloon pump use at the University of Miami/Jackson Memorial Medical Center over the past 21 years identified 2 cases where a balloon was found to be entrapped. The balloon catheters had been in place for approximately 10 days when this complication occurred. The retained balloons were torn, filled with clotted blood, and impacted in the vasculature. In our first case, forceful removal of the intraaortic balloon was complicated by unintentional extraction of the external iliac and common femoral arteries. In the second case, clot within the balloon was dissolved with tissue plasminogen activator injected into the drive lumen of the catheter before removal. The prevention and management of this rare but serious complication of intraaortic balloon pumping is reviewed.

Entities:  

Mesh:

Year:  1993        PMID: 8347025     DOI: 10.1016/0003-4975(93)91180-u

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  6 in total

1.  Vascular trauma in civilian practice.

Authors:  S Clark
Journal:  Ann R Coll Surg Engl       Date:  1996-03       Impact factor: 1.891

2.  Management of Intra-Aortic Balloon Pump Rupture and Entrapment.

Authors:  Madhu Bhamidipaty; Barend Mees; Timothy Wagner
Journal:  Aorta (Stamford)       Date:  2016-04-01

Review 3.  Management of intra-aortic balloon pump entrapment: a case report and review of the literature.

Authors:  Gerard J Fitzmaurice; Anton Collins; Haralambos Parissis
Journal:  Tex Heart Inst J       Date:  2012

4.  Limb salvage following iatrogenic arterial injury: complications of cardiac support using intra-aortic balloon pumps.

Authors:  Adam Hunt; Owen Waldin; Hani Slim; Ranjeet Brar
Journal:  BMJ Case Rep       Date:  2016-06-06

Review 5.  Retained intraaortic balloon. Case report and review of the literature.

Authors:  A M Grande; L Martinelli; A Graffigna; M Viganò
Journal:  Tex Heart Inst J       Date:  1995

6.  Intra-Aortic Balloon Pump Entrapment in a Nonatherosclerotic Common Iliac Artery.

Authors:  Feridoun Sabzi; Abdol Hamid Zokaei; Abdol Rasoul Moloudi; Samsam Dabiri
Journal:  Open J Cardiovasc Surg       Date:  2011-09-19
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.