Literature DB >> 8326303

The 3-6-year results of a modular noncemented low-bending stiffness hip implant. A preliminary study.

H U Cameron1.   

Abstract

A review of the 3-6-year clinical and radiographic results of a modular noncemented stem has been carried out. This is a three-part stem consisting of a head, stem, and proximal sleeve. The stems are split distally like a clothespin in the coronal plane to reduce bending stiffness. No mechanical failures have occurred, including no bead separation from the single-layer porous-coated taper lock sleeves. Of the 48 cases, 13 had smooth stems distally while the rest were fluted. Five of the fluted stems were solid, that is, nonsplit. Clinical results are 93.7% excellent, 4.2% good, and 2.1% poor with the Harris hip rating. There have been no revisions. Thigh pain incidence with the distal split stem was 4.4%. Radiographic follow-up evaluations revealed nonprogressive radiolucency around one of the ingrowth sleeves. When inserted in varus the smooth stems developed some radiolucency around the distal end. Fluting of the stem appeared to prevent this.

Entities:  

Mesh:

Year:  1993        PMID: 8326303     DOI: 10.1016/s0883-5403(06)80084-7

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  7 in total

1.  Spontaneous fracture of diaphyseal stem of S-ROM femoral prosthesis.

Authors:  Simon Pearce; Amir-Reza Jenabzadeh; William L Walter; Ronald Mark Gillies
Journal:  BMJ Case Rep       Date:  2014-05-20

2.  Revision hip arthroplasty using a cementless modular tapered stem.

Authors:  Christophe Pattyn; Alexander Mulliez; René Verdonk; Emmanuel Audenaert
Journal:  Int Orthop       Date:  2011-06-24       Impact factor: 3.075

3.  Modular femoral sleeve and stem implant provides long-term total hip survivorship.

Authors:  David Le; Karen Smith; Dylan Tanzer; Michael Tanzer
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

4.  Late remodeling around a proximally HA-coated tapered titanium femoral component.

Authors:  William N Capello; James A D'Antonio; Rudolph G Geesink; Judy R Feinberg; Marybeth Naughton
Journal:  Clin Orthop Relat Res       Date:  2008-10-11       Impact factor: 4.176

5.  Total hip arthroplasty using S-ROM prosthesis for dysplastic hip.

Authors:  Joon Soon Kang; Kyoung Ho Moon; Ryuh Sup Kim; Seung Rim Park; Jung Sun Lee; Sang Hyun Shin
Journal:  Yonsei Med J       Date:  2011-07       Impact factor: 2.759

6.  Load Sharing in the Femur Using Strut Allografts: A Biomechanical Study.

Authors:  Frances E Sharpe; Kipling P Sharpe; Colin P McCarty; Edward Ebramzadeh
Journal:  Arthroplast Today       Date:  2022-04-12

Review 7.  Modular Stems: Advantages and Current Role in Primary Total Hip Arthroplasty.

Authors:  Chan-Woo Park; Seung-Jae Lim; Youn-Soo Park
Journal:  Hip Pelvis       Date:  2018-09-04
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.