Performance characteristics of the HemoCue B-Glucose analyzer using whole-blood samples.

E M Voss1, G S Cembrowski.   

Abstract

We evaluated the HemoCue B-Glucose (HemoCue Inc, Mission Viejo, Calif) analyzer for accuracy, precision, linearity, and recovery. One hundred eighteen capillary whole-blood samples were analyzed in duplicate on the HemoCue B-Glucose and the YSI 2300 STAT Glucose/L-Lactate (Yellow Springs [Ohio] Instruments) analyzers; corresponding plasma glucose levels were measured in duplicate on the Roche Cobas MIRA (Roche Diagnostic Systems, Nutley, NJ) analyzer. Plasma glucose levels were converted to whole-blood equivalent glucose levels by using a factor of 1.11. The following regression equations were obtained: HemoCue = 1.02 (YSI) + 0.19, Sy/x = 0.52, r2 = .984; and HemoCue = 0.98 (whole-blood equivalent glucose levels) + 0.26, Sy/x = 0.55, r2 = .982. Within-run coefficients of variation were 4.0%, 3.5%, 2.2%, and 1.0% at glucose concentrations of 3.9, 5.4, 8.7, and 17.1 mmol/L (71, 97, 156, and 308 mg/dL), respectively. Between-run imprecision and total imprecision using lyopholized materials with three lot numbers of cuvettes were 4.2% and 5.6% at 2.1 mmol/L (37 mg/dL) and 2.4% and 2.7% at 5.2 mmol/L (95 mg/dL), respectively. The HemoCue B-Glucose analyzer displayed linearity between 0 and 22.2 mmol/L (0 and 400 mg/dL), and the percent recovery averaged 98.7% +/- 4.5% (mean +/- SD). The HemoCue B-Glucose analyzer is a rapid, simple, and reliable method for determinations of whole-blood glucose levels.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8323435

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  6 in total

1.  Population-based stepwise screening for unrecognised Type 2 diabetes is ineffective in general practice despite reliable algorithms.

Authors:  J O Christensen; A Sandbaek; T Lauritzen; K Borch-Johnsen
Journal:  Diabetologia       Date:  2004-09-08       Impact factor: 10.122

2.  Blood volume determination in obese and normal-weight gravidas: the hydroxyethyl starch method.

Authors:  Laura K Vricella; Judette M Louis; Edward Chien; Brian M Mercer
Journal:  Am J Obstet Gynecol       Date:  2015-05-14       Impact factor: 8.661

3.  Evaluation of the VIA Blood Chemistry Monitor for Glucose in Healthy and Diabetic Volunteers.

Authors:  Arjunan Ganesh; Brian Hipszer; Navdeep Loomba; Barbara Simon; Marc C Torjman; Jeffrey Joseph
Journal:  J Diabetes Sci Technol       Date:  2008-03

Review 4.  Interpretation of EQA results and EQA-based trouble shooting.

Authors:  Gunn Berit Berge Kristensen; Piet Meijer
Journal:  Biochem Med (Zagreb)       Date:  2017-02-15       Impact factor: 2.313

5.  Most Women with Previous Gestational Diabetes Mellitus Have Impaired Glucose Metabolism after a Decade.

Authors:  Jeanette Wahlberg; Bertil Ekman; Hans J Arnqvist
Journal:  Int J Mol Sci       Date:  2018-11-23       Impact factor: 5.923

6.  Evaluation of a Novel Noninvasive Blood Glucose Monitor Based on Mid-Infrared Quantum Cascade Laser Technology and Photothermal Detection.

Authors:  Thorsten Lubinski; Bartosz Plotka; Sergius Janik; Luca Canini; Werner Mäntele
Journal:  J Diabetes Sci Technol       Date:  2020-07-05
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.