Intrinsic PEEP determined by static pressure-volume curves--application of a novel automated occlusion method.

M Sydow1, H Burchardi, J Zinserling, T A Crozier, T Denecke, S Zielmann.   

Abstract

OBJECTIVE: Evaluation of new computer-controlled occlusion procedure for determination of intrinsic PEEP in mechanically ventilated patients and comparison with the standard end-expiratory occlusion method.
DESIGN: Prospective controlled study.
SETTING: Intensive care unit of a university hospital. PATIENTS: 16 patients with acute respiratory failure of different degree and etiology. All patients were mechanically ventilated, heavily sedated and muscle paralyzed (non-depolarising relaxants). The type of ventilator, the inspiration/expiration ratio, FIO2 and PEEP were selected by the attending clinicians according to the patients' need and independently from the study.
INTERVENTIONS: Static compliance of the respiratory system (Cstat) was determined at varying external end-expiratory pressure settings: ZEEP (= ambient pressure), PEEP of 5 cmH2O and 10 cmH2O. All other ventilator settings were kept constant during the entire procedure. MEASUREMENTS AND
RESULTS: A computer-controlled occlusion method (SCASS) was used for determination of Cstat. Intrinsic PEEP was determined by SCASS as the extrapolated zero-volume intercept of the regression line of multiple pressure/volume data pairs (PEEPSCASSinspir and PEEPSCASSexpir). Directly thereafter intrinsic PEEP in this particular ventilatory setting was determined by end-expiratory occlusions (PEPPEEO). The intrinsic PEEP values of the different methods were nearly identical with a significant correlation (p < 0.0001). Mean values +/- SD: PEEPSCASSinspir 7.1 +/- 4.3 cmH2O; PEEPSCASSexpir 7.1 +/- 4.5 cmH2O; PEEPEEO 7.1 +/- 4.2 cmH2O.
CONCLUSION: Since no significant difference between PEEPi values measured by the inspiratory and expiratory occlusion method (SCASS) was seen, this indicates that no alveolar recruitment occurred during the respiratory cycle. This study demonstrates that the automated occlusion method for measuring Cstat system can also be used with high accuracy for determination of intrinsic PEEP in mechanically ventilated patients.

Entities:  

Mesh:

Year:  1993        PMID: 8315125     DOI: 10.1007/bf01720534

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  19 in total

1.  Regional distribution of pulmonary ventilation and perfusion in elderly subjects.

Authors:  J Holland; J Milic-Emili; P T Macklem; D V Bates
Journal:  J Clin Invest       Date:  1968-01       Impact factor: 14.808

2.  Causes of error of respiratory pressure-volume curves in paralyzed subjects.

Authors:  J Dall'ava-Santucci; A Armaganidis; F Brunet; J F Dhainaut; G L Chelucci; J F Monsallier; A Lockhart
Journal:  J Appl Physiol (1985)       Date:  1988-01

3.  Noninvasive determination of respiratory system mechanics during mechanical ventilation for acute respiratory failure.

Authors:  S B Gottfried; A Rossi; B D Higgs; P M Calverley; L Zocchi; C Bozic; J Milic-Emili
Journal:  Am Rev Respir Dis       Date:  1985-03

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Respiratory compliance and resistance in mechanically ventilated patients with acute respiratory failure.

Authors:  M Bernasconi; Y Ploysongsang; S B Gottfried; J Milic-Emili; A Rossi
Journal:  Intensive Care Med       Date:  1988       Impact factor: 17.440

6.  Occult positive end-expiratory pressure in mechanically ventilated patients with airflow obstruction: the auto-PEEP effect.

Authors:  P E Pepe; J J Marini
Journal:  Am Rev Respir Dis       Date:  1982-07

7.  The role of total static lung compliance in the management of severe ARDS unresponsive to conventional treatment.

Authors:  L Gattinoni; A Pesenti; M L Caspani; A Pelizzola; D Mascheroni; R Marcolin; G Iapichino; M Langer; A Agostoni; T Kolobow
Journal:  Intensive Care Med       Date:  1984       Impact factor: 17.440

8.  Dynamic hyperinflation and ventilator dependence in chronic obstructive pulmonary disease.

Authors:  W R Kimball; D E Leith; A G Robins
Journal:  Am Rev Respir Dis       Date:  1982-12

9.  Improved determination of static compliance by automated single volume steps in ventilated patients.

Authors:  M Sydow; H Burchardi; J Zinserling; H Ische; T A Crozier; W Weyland
Journal:  Intensive Care Med       Date:  1991       Impact factor: 17.440

10.  Atelectasis during anaesthesia and in the postoperative period.

Authors:  A Strandberg; L Tokics; B Brismar; H Lundquist; G Hedenstierna
Journal:  Acta Anaesthesiol Scand       Date:  1986-02       Impact factor: 2.105

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.