Silicone breast implant rupture: comparison between three-point Dixon and fast spin-echo MR imaging.

D P Gorczyca1, E Schneider, N D DeBruhl, T K Foo, C Y Ahn, J W Sayre, W W Shaw, L W Bassett.   

Abstract

OBJECTIVE: This study was designed to compare the three-point Dixon technique with our present MR protocol incorporating T2-weighted fast spin echo and fast spin echo with water suppression to detect ruptured silicone breast implants. SUBJECTS AND METHODS: Eighty-two symptomatic women with silicone breast implants were examined with both the three-point Dixon technique and fast spin-echo MR sequences. Of these patients, 41 had surgery to remove their implants. Four radiologists reviewed the images from only those patients who had surgery and graded each for rupture by using a scale of 1-5. Receiver-operating-characteristic analysis was performed.
RESULTS: Of 81 implants removed, 18 were ruptured. Silicone implant ruptures were identified more frequently on the fast spin-echo sequence than on the three-point Dixon sequence, with areas under the ROC curves of .95 and .84, respectively. Although the difference was not statistically significant, the sensitivity for detecting silicone implant rupture was 89% for the fast spin-echo sequence and 61% for the three-point Dixon sequence. The specificity was 97% for both sequences.
CONCLUSION: Silicone implant ruptures were detected more frequently with fast spin-echo MR sequences than with the three-point Dixon technique, although the difference was not significant. The greater spatial resolution used for the fast spin-echo sequence partially accounts for the difference in detection of implant ruptures in this study.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8310916     DOI: 10.2214/ajr.162.2.8310916

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  3 in total

1. 

Authors:  H Tristant
Journal:  Arch Gynecol Obstet       Date:  1995-12       Impact factor: 2.344

Review 2.  The effect of study design biases on the diagnostic accuracy of magnetic resonance imaging for detecting silicone breast implant ruptures: a meta-analysis.

Authors:  Jae W Song; Hyungjin Myra Kim; Lillian T Bellfi; Kevin C Chung
Journal:  Plast Reconstr Surg       Date:  2011-03       Impact factor: 4.730

3.  [Breast cancer. Sonography and magnetic resonance mammography].

Authors:  S Delorme
Journal:  Radiologe       Date:  2004-06       Impact factor: 0.635

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.