Optimal sampling interval and edge detection algorithm for measurement of blood vessel diameter on a cineangiogram.

C Uyama1, Y Kita, S Matsusita.   

Abstract

RATIONALE AND
OBJECTIVES: The authors studied the effects of image magnification and the type of edge detection filter on the precision of measuring blood vessel diameters on coronary angiograms.
METHOD: A blood vessel phantom containing five channels of various diameters filled with contrast medium was filmed. The magnifications examined were x4, x10, and x20, and the edge detection filters used were the first derivative, second derivative, composite, and entropy filters. The regression line y = a+bx was introduced, where x represents the nominal diameter and y the diameter measured by the edge-detection filters. Determination criteria were the offset, slope, and residual variance from the regression equation.
RESULTS: The best value among three criteria was obtained with a magnification of x10 or x20. The entropy filter gave the best value of slope and residual variance. The composite filter gave the best value of offset.
CONCLUSION: The most precise measurement is obtained when the entropy filter and a magnification of x10 or x20 are applied.

Mesh:

Year:  1993        PMID: 8307716     DOI: 10.1097/00004424-199312000-00008

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  2 in total

1.  Image correlation algorithm for measuring lymphocyte velocity and diameter changes in contracting microlymphatics.

Authors:  J Brandon Dixon; Anatoliy A Gashev; David C Zawieja; James E Moore; Gerard L Coté
Journal:  Ann Biomed Eng       Date:  2006-12-07       Impact factor: 3.934

2.  Angiographic film subtraction using a laser digitizer and computer processing.

Authors:  J M Boone; N M Corrigan; S T Hecht; D P Link
Journal:  J Digit Imaging       Date:  1998-11       Impact factor: 4.056

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.