Exposure reduction in cephalography with a digital photostimulable phosphor imaging system.

K Seki1, T Okano.   

Abstract

The effect on the determination of the cephalometric landmarks of exposure reduction using computed radiography with a photostimulable storage phosphor was evaluated. Forty patients attending for regular review were divided into four groups of 10. The first group was exposed to the same amount of radiation as that required for a conventional calcium tungstate screen-film combination. The other three groups were exposed to approximately 50, 20 and 10% respectively of the radiation dose of the first group. Ten dentists determined the landmarks. The coordinates of the points were recorded to measure the consistency of landmark identification. It was found that in comparison with the conventional screen-film system, a substantial reduction of exposure could be achieved with computed radiography for cephalography without clinical detriment.

Entities:  

Mesh:

Year:  1993        PMID: 8299830     DOI: 10.1259/dmfr.22.3.8299830

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  2 in total

1.  Prospective study on the reproducibility of cephalometric landmarks on conventional and digital lateral headfilms.

Authors:  K Hagemann; D Vollmer; T Niegel; U Ehmer; I Reuter
Journal:  J Orofac Orthop       Date:  2000       Impact factor: 1.938

2.  Assessing the Reliability of Digitalized Cephalometric Analysis in Comparison with Manual Cephalometric Analysis.

Authors:  Mohammed Umar Farooq; Mohd Asadullah Khan; Shahid Imran; Ayesha Sameera; Arshad Qureshi; Syed Afroz Ahmed; Sujan Kumar; Mohd Aziz Ur Rahman
Journal:  J Clin Diagn Res       Date:  2016-10-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.