Measurement of respiratory mechanics using the Puritan-Bennett 7200a ventilator.

D Chartrand1, B Dionne, C Jodoin, M Lorange, A Lapointe.   

Abstract

This study was designed in order to validate the respiratory mechanical variables measured by the Puritan-Bennett 7200a ventilator equipped with the 30/40 module. Two ventilators were connected to a lung model and submitted to several breathing patterns by modifying the respiratory rate, the tidal volume, the inspiratory flow-rate and the model resistance. The inspiratory flow-rate (V), tidal volume (VT), peak inspiratory pressure (Pmax), plateau pressure (Pplat) and PEEP measured by the ventilators were compared with the same variables measured at the connection between the breathing circuit and the lung model. The compliance (C30/40) and the resistance (R30/40) calculated by the 30/40 module were compared with those calculated by using the variables measured by the reference equipment. Both ventilators made a constant underestimation of V by 2.8 and 3.7 L.min-1, respectively. The VT was measured with a mean error of less than 10 ml but did not reflect the preselected values in the presence of an intrinsic PEEP. The Pplat was overestimated by 7 and 10%, respectively. The same calibration error was observed with Pmax which was also affected by a pressure gradient due to the resistance of the breathing circuit. Even in the absence of intrinsic PEEP, C30/40 presented an error due to the combination of the measurement errors on VT, Pplat and PEEP. Finally, R30/40 presented a high percentage of error due to the combination of the measurement errors on V, Pmax, and Pplat, and to a sporadic aberrant selection of V. Due to these numerous sources of error, the two ventilators studied did not give reliable estimates of resistance and compliance.(ABSTRACT TRUNCATED AT 250 WORDS)

Mesh:

Substances:

Year:  1993        PMID: 8269571     DOI: 10.1007/BF03009481

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  9 in total

1.  THE MECHANICS OF THE RESPIRATORY SYSTEM DURING ANESTHESIA. THE EFFECTS OF ATROPINE AND CARBON DIOXIDE.

Authors:  H F DON; J G ROBSON
Journal:  Anesthesiology       Date:  1965 Mar-Apr       Impact factor: 7.892

2.  Estimation of time-varying respiratory mechanical parameters by recursive least squares.

Authors:  A M Lauzon; J H Bates
Journal:  J Appl Physiol (1985)       Date:  1991-09

3.  Automated measurement of respiratory mechanics in anaesthetized ventilated patients.

Authors:  D Benhamou; A M Lorino; H Lorino; F Zerah; A Harf
Journal:  Bull Eur Physiopathol Respir       Date:  1987 Sep-Oct

4.  A computerized method for measuring respiratory mechanics during mechanical ventilation.

Authors:  A M Lorino; D Benhamou; H Lorino; A Harf
Journal:  Bull Eur Physiopathol Respir       Date:  1986 Jan-Feb

5.  A comparison of some methods for measuring total respiratory resistance.

Authors:  N A Bergman; C L Waltemath
Journal:  J Appl Physiol       Date:  1974-01       Impact factor: 3.531

6.  Analysis of the behavior of the respiratory system with constant inspiratory flow.

Authors:  J H Bates; A Rossi; J Milic-Emili
Journal:  J Appl Physiol (1985)       Date:  1985-06

7.  Respiratory mechanics in mechanically ventilated patients with respiratory failure.

Authors:  A Rossi; S B Gottfried; B D Higgs; L Zocchi; A Grassino; J Milic-Emili
Journal:  J Appl Physiol (1985)       Date:  1985-06

8.  [Automated monitoring of respiratory parameters in the anesthetized patient in mechanical respiration].

Authors:  A M Lorino; I Cigarini; M Bénichou; F Bonnet; H Lorino; A Harf
Journal:  Ann Fr Anesth Reanim       Date:  1988

9.  A theoretical analysis of interrupter technique for measuring respiratory mechanics.

Authors:  J H Bates; P Baconnier; J Milic-Emili
Journal:  J Appl Physiol (1985)       Date:  1988-05
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.