Literature DB >> 8269491

Total aortic occlusion caused by sustained balloon inflation: a previously unreported complication of intraaortic balloon counterpulsation.

M A Ferrell1, M Doherty, R M Zusman, I S Nash.   

Abstract

A case of acute lower-body ischemia 2 days following intraaortic balloon pump insertion is reported. Fluoroscopy revealed persistent balloon inflation throughout the cardiac cycle with distal aortic occlusion. Attempts to manually deflate the balloon were unsuccessful until a guidewire was advanced through the gas-exchange lumen into the body of the balloon. The balloon catheter was removed without clinical sequelae other than transient oliguria and an asymptomatic increase in creatinine phosphokinase (CPK). This is a previously unreported complication of intraaortic balloon counterpulsation.

Entities:  

Mesh:

Year:  1993        PMID: 8269491     DOI: 10.1002/ccd.1810300306

Source DB:  PubMed          Journal:  Cathet Cardiovasc Diagn        ISSN: 0098-6569


  1 in total

1.  Limb salvage following iatrogenic arterial injury: complications of cardiac support using intra-aortic balloon pumps.

Authors:  Adam Hunt; Owen Waldin; Hani Slim; Ranjeet Brar
Journal:  BMJ Case Rep       Date:  2016-06-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.