| Literature DB >> 8269491 |
M A Ferrell1, M Doherty, R M Zusman, I S Nash.
Abstract
A case of acute lower-body ischemia 2 days following intraaortic balloon pump insertion is reported. Fluoroscopy revealed persistent balloon inflation throughout the cardiac cycle with distal aortic occlusion. Attempts to manually deflate the balloon were unsuccessful until a guidewire was advanced through the gas-exchange lumen into the body of the balloon. The balloon catheter was removed without clinical sequelae other than transient oliguria and an asymptomatic increase in creatinine phosphokinase (CPK). This is a previously unreported complication of intraaortic balloon counterpulsation.Entities:
Mesh:
Year: 1993 PMID: 8269491 DOI: 10.1002/ccd.1810300306
Source DB: PubMed Journal: Cathet Cardiovasc Diagn ISSN: 0098-6569