Abstract
A case is presented in which a curved needle became lost during extracorporeal suturing in laparoscopic surgery. Maneuvers to find the needle were fruitless until an intraoperative x-ray showed it was stuck in the flapper mechanism of a laparoscopic cannula. Caution is given against the use of "pop-off" needles during extracorporeal suturing and an x-ray of cannulae is suggested when a lost needle cannot be found using routine maneuvers.Mesh:
Year: 1993 PMID: 8268519 DOI: 10.1089/lps.1993.3.425