Literature DB >> 8267991

The value of the stentless biovalve prosthesis. An experimental study.

M G Hazekamp1, Y A Goffin, H A Huysmans.   

Abstract

Clinical use of the stentless bioprosthesis has not yet been accepted world-wide. Experimental studies are scarce. In a study in growing pigs, 23 aortic valve replacements were performed (7 stented bioprostheses (STB), 10 stentless bioprostheses (SLB) and 6 cryopreserved homografts (CAH)). Valves were studied macroscopically and microscopically after explantation. Five animals died between 1.5 and 4 months after implantation. Two STBs showed extreme calcific degeneration. A third STB animal died suddenly: this STB showed only minimal leaflet calcification. One SLB was stenosed with a large fibrin deposit in one cusp, a second SLB showed slight cusp calcification and three were paracommissural tears in one cusp. In all other animals the valves were explanted after 5 to 6.5 months. All STBs showed severe calcific degeneration. Five SLBs showed mild calcific degeneration, while three others were unaltered after 6 months. All CAHs were free of calcific degeneration, three were perfect, two had a tear in a commisure and another was prolapsed with a fibrin degeneration. The speed and extent of valve degeneration were less than in the STB group. The stentless design is an important contribution which may result in a higher durability of bioprosthetic valves.

Entities:  

Mesh:

Year:  1993        PMID: 8267991     DOI: 10.1016/1010-7940(93)90048-g

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  1 in total

1.  Dynamic obstruction, an unusual complication after aortic valve replacement with a stentless porcine valve.

Authors:  R P van Roosmalen; L H Baur; J Braun; D E Atsma; M G Hazekamp; E E van der Wall; H A Huysmans
Journal:  Int J Card Imaging       Date:  1999-06
  1 in total