Literature DB >> 8264209

[The Array multifocal lens--functional principle and clinical results].

D Eisenmann1, K W Jacobi.   

Abstract

BACKGROUND: Common refractive bifocal IOLs feature 2 or 3 spherical zones with different refraction; the design of the Array multifocal IOL consists of 5 concentric zones, each of them providing a progressive near addition with multiple foci by means of an aspherical surface.
METHODS: As a part of a prospective study, distance and near visual acuity, contrast sensitivity, glare and depth of field were examined in 15 patients with the Array lens and a monofocal control group 12 months after implantation. Subjective assessment of the optical quality was performed by detailed asking for optical symptoms. Results were retrospectively compared with the one year follow-up of 15 patients with diffractive bifocal IOLs.
RESULTS: No difference was found for best corrected far and near visual acuity and sensitivity for high contrasts. Glare and sensitivity for low contrasts of the monofocal IOL was significantly superior to both multifocal models. Array IOL and diffractive IOL achieved a near acuity of Nieden 2-1 with only distance correction; defocus curves revealed an increase in the depth of field, but with a superiority of the Array lens in intermediate imaging. Optical symptoms as halos were much more frequent in patients with the diffractive IOL than in the Array group.
CONCLUSIONS: Functional results of the Array multifocal IOL seem to be comparable to those of the diffractive bifocal IOL, but with the advantage of an improved intermediate imaging and a reduction of optical side effects.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8264209     DOI: 10.1055/s-2008-1045666

Source DB:  PubMed          Journal:  Klin Monbl Augenheilkd        ISSN: 0023-2165            Impact factor:   0.700


  3 in total

Review 1.  [Apodized diffractive optic. New concept in multifocal lens technology].

Authors:  T Kohnen; V Derhartunian
Journal:  Ophthalmologe       Date:  2007-10       Impact factor: 1.059

2.  Halo size under distance and near conditions in refractive multifocal intraocular lenses.

Authors:  S Pieh; B Lackner; G Hanselmayer; R Zöhrer; M Sticker; H Weghaupt; A Fercher; C Skorpik
Journal:  Br J Ophthalmol       Date:  2001-07       Impact factor: 4.638

3.  Contrast sensitivity and glare disability by halogen light after monofocal and multifocal lens implantation.

Authors:  S Schmitz; H B Dick; F Krummenauer; O Schwenn; R Krist
Journal:  Br J Ophthalmol       Date:  2000-10       Impact factor: 4.638

  3 in total