Literature DB >> 8245339

Atrial sensing performance of the single-lead VDD pacemaker during exercise.

P Varriale1, B E Chryssos.   

Abstract

OBJECTIVES: The purpose of this study was to examine the atrial sensing performance of the single-lead VDD pacing system during exercise and concomitant changes in the amplitude of the atrial electrogram.
BACKGROUND: Studies of conventional dual-chamber pacing have demonstrated an overall reduction in the atrial signal amplitude and a variable incidence of atrial undersensing during vigorous exercise.
METHODS: The telemetered atrial electrogram and simultaneous surface electrocardiogram (ECG) were continuously recorded in 12 patients (mean age 70.8 years) with an implanted single-lead VDD pacing system during treadmill stress testing. The atrial signal amplitude was measured at peak exercise, and the patients were monitored for maintained atrial synchronized ventricular pacing during the entire exercise and recovery period.
RESULTS: The atrial electrographic voltage decreased an average of 19.5% (p < 0.05) during peak exercise, and the reduced P wave amplitude ranged from 6.9% to 59.4% of the preexercise values in 8 of 12 patients. Three patients showed a modest increase in atrial signal amplitude, and one patient had no change. The telemetered electrogram displayed persistent and intact atrial synchronous ventricular pacing throughout the study period.
CONCLUSIONS: Despite relatively low atrial signal amplitudes at rest and further decreases during exercise, the single-lead VDD pacemaker maintains reliable atrial tracking and ventricular pacing during vigorous exercise.

Entities:  

Mesh:

Year:  1993        PMID: 8245339     DOI: 10.1016/0735-1097(93)90769-w

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  4 in total

1.  Predictors of loss of atrioventricular synchrony in single lead VDD pacing.

Authors:  P Hunziker; P Buser; M Pfisterer; F Burkart; S Osswald
Journal:  Heart       Date:  1998-10       Impact factor: 5.994

Review 2.  The current status of single lead dual chamber sensing and pacing.

Authors:  H F Tse; C P Lau
Journal:  J Interv Card Electrophysiol       Date:  1998-09       Impact factor: 1.900

3.  Intracardiac atrial signal amplitude in congenital and acquired complete heart block.

Authors:  Mohammad Reza Samiei Nasab; Mohammad Reza Dehghani; Mehrdad Taherioun; Alireza Rostamzadeh
Journal:  Indian Pacing Electrophysiol J       Date:  2010-10-31

Review 4.  Cardiac implantable devices during exercise: Normal function and troubleshooting.

Authors:  Oswaldo J Gutiérrez
Journal:  J Arrhythm       Date:  2021-03-22
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.