Repeated use of ablation catheters: a prospective study.

B Avitall1, M Khan, D Krum, M Jazayeri, J Hare.   

Abstract

OBJECTIVES: The objective of this study was to provide insight into the time course of electrical, physical and mechanical changes in ablation catheters after each use that may affect the safety and efficacy of the ablation procedure.
BACKGROUND: An increasing number of institutions are reusing deflectable ablation catheters. At present, there are no data concerning the safety of reusing ablation catheters.
METHODS: Over a period of 1 year, 69 Webster/Mansfield deflectable catheters used in 336 ablation procedures were prospectively studied. An additional 18 new catheters were tested after multiple sterilizations only. The catheters were evaluated for electrical and physical integrity and mechanical capabilities. These include deflection at room temperature and 37 degrees C, shaft compression and buckling during deflection, tip craters, torquing ability, glue separation and tip attachment using a stereoscope at x30 magnification and electrical resistance for each electrode. After each use, the catheters were gas-sterilized with ethylene oxide.
RESULTS: The most common reasons for catheter rejection were tip electrode glue separation after 4.3 +/- 4.3 uses and loss of deflection after 5.0 +/- 3.3 uses. Electrical discontinuity between the catheter handle and electrodes was observed after 10.0 +/- 3.7 uses. There was no significant decrease in catheter torquing ability with repeated use. In this study the total estimated savings was $128,133, which includes the cost of catheter reprocessing. The reuse of Webster/Mansfield ablation catheters has not resulted in any major catheter failure or any major adverse clinical complications.
CONCLUSIONS: On the basis of these observations, we believe that the Webster/Mansfield catheter can be reused an average of five times. It is strongly recommended that after each use catheters be carefully examined under appropriate magnification (x30) and that special attention be given to the ablation tip electrode. The catheters should also be tested for deflection and electrical integrity.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8227793     DOI: 10.1016/0735-1097(93)90544-b

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  7 in total

1.  In vitro performance characteristics of reused ablation catheters.

Authors:  Igor Ayzman; Samer R Dibs; Jeffrey Goldberger; Rod Passman; Alan Kadish
Journal:  J Interv Card Electrophysiol       Date:  2002-08       Impact factor: 1.900

2.  Comparison of performance characteristics between new and reprocessed electrophysiology catheters.

Authors:  Bruce R Lester; Alan A Alexander; Keith Miller; Nicole P Boser; Brian F Sullivan; Gregory G Brucker
Journal:  J Interv Card Electrophysiol       Date:  2007-01-17       Impact factor: 1.900

Review 3.  Cardiac arrhythmias in low- and middle-income countries.

Authors:  Philasande Mkoko; Ehete Bahiru; Olujimi A Ajijola; Aime Bonny; Ashley Chin
Journal:  Cardiovasc Diagn Ther       Date:  2020-04

Review 4.  Guidance on reuse of cardio-vascular catheters and devices in India: A consensus document.

Authors:  Aditya Kapoor; Amit Vora; Gita Nataraj; Sundeep Mishra; Prafulla Kerkar; C N Manjunath
Journal:  Indian Heart J       Date:  2017-04-13

Review 5.  Reprocessing and reuse of urological armamentarium: How correct are we!

Authors:  Krutik Vipulbhai Raval; Rajeev Chaudhari; Shahil Rameshbhai Khant; Omkar Joglekar; Dipen Patel
Journal:  Urol Ann       Date:  2017 Apr-Jun

6.  Reprocessing of Medical Products in Electrophysiology.

Authors:  Ricardo Ryoshim Kuniyoshi; Eduardo Back Sternick; Elenir Nadalin; Denise Tessariol Hachul
Journal:  Arq Bras Cardiol       Date:  2017-02       Impact factor: 2.000

7.  Catheter steering in interventional cardiology: Mechanical analysis and novel solution.

Authors:  Awaz Ali; Aimee Sakes; Ewout A Arkenbout; Paul Henselmans; Remi van Starkenburg; Tamas Szili-Torok; Paul Breedveld
Journal:  Proc Inst Mech Eng H       Date:  2019-10-03       Impact factor: 1.617

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.