In vitro performance characteristics of reused ablation catheters.

Igor Ayzman1, Samer R Dibs, Jeffrey Goldberger, Rod Passman, Alan Kadish.   

Abstract

BACKGROUND: Prior studies have found that there is a widespread practice of catheter reprocessing in cardiac electrophysiology laboratories. Effects of reprocessing of ablation catheters on temperature sensing and mechanical deflection are not fully known.
METHODS: Twenty-four new and used ablation catheters were studied. Deviation of temperature sensing by catheters from the temperature of a heated saline bath was measured. The angle of deflection of digitally scanned catheters at 75% and 100% handle deflection was also measured. New and used catheters were compared with respect to their temperature sensing accuracy and deflection characteristics.
RESULTS: Overall, there was 0.7 +/- 0.1 degrees C (mean +/- standard error) deviation of the sensed temperature from the bath temperature, with no significant difference between new and used catheters. Similarly, there was no significant difference in the angle of deflection between new (66.7 degrees +/- 6.2 degrees and 24.3 degrees +/- 6.8 degrees at 75% and 100% deflections, respectively) and used (59.6 degrees +/- 5.6 degrees and 28.7 degrees +/- 9.9 degrees at 75% and 100% deflections, respectively) catheters. The difference in the angle of deflection between matched new and used catheters was 18.9 degrees +/- 4.2 degrees and 10.9 degrees +/- 2.4 degrees at 75% and 100% deflections, respectively, with a relatively broad range (5.0 degrees -35.6 degrees and 0.4 degrees -19.0 degrees at 75% and 100% deflections, respectively).
CONCLUSIONS: This study found no significant overall difference in temperature sensing accuracy and deflection angle of new and used ablation catheters. Nevertheless, individual differences in deflection characteristics between new and used catheters are occasionally seen and warrant screening of reprocessed catheters prior to their reuse.

Entities:  

Mesh:

Year:  2002        PMID: 12391420     DOI: 10.1023/a:1020820200112

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  21 in total

1.  Risk of infection by reprocessed and resterilized virus-contaminated catheters; an in-vitro study.

Authors:  D S Luijt; J Schirm; P H Savelkoul; A Hoekstra
Journal:  Eur Heart J       Date:  2001-03       Impact factor: 29.983

2.  Reprocessing of electrophysiology catheters: clinical studies, regulations, and recommendations. A report of the NASPE Task Force on Reprocessing of Electrophysiological Catheters. North American Society of Pacing and Electrophysiology.

Authors:  B D Lindsay; S P Kutalek; D S Cannom; S C Hammill; G V Naccarelli
Journal:  Pacing Clin Electrophysiol       Date:  2001-08       Impact factor: 1.976

3.  The safety of reusing ablation catheters with temperature control and the need for a validation protocol and guidelines for reprocessing.

Authors:  C Blomström-Lundqvist
Journal:  Pacing Clin Electrophysiol       Date:  1998-12       Impact factor: 1.976

4.  Catheter reuse: boon or boondoggle?

Authors:  D M Gilligan; M A Wood; K A Ellenbogen
Journal:  Pacing Clin Electrophysiol       Date:  1998-12       Impact factor: 1.976

5.  Reuse of disposables: let's not embrace waste.

Authors:  Z G Turi
Journal:  Cathet Cardiovasc Diagn       Date:  1996-06

6.  Cost efficacy modeling of catheter reuse for percutaneous transluminal coronary angioplasty.

Authors:  M D Lesh
Journal:  J Am Coll Cardiol       Date:  1996-11-15       Impact factor: 24.094

7.  Reuse of pacing catheters: a survey of safety and efficacy.

Authors:  S O'Donoghue; E V Platia
Journal:  Pacing Clin Electrophysiol       Date:  1988-09       Impact factor: 1.976

8.  Ethylene oxide on electrophysiology catheters following resterilization: implications for catheter reuse.

Authors:  M Ferrell; C E Wolf; K A Ellenbogen; M A Wood; H F Clemo; D M Gilligan
Journal:  Am J Cardiol       Date:  1997-12-15       Impact factor: 2.778

Review 9.  Reuse of angioplasty catheters and risk of Creutzfeldt-Jakob disease.

Authors:  B Fagih; M J Eisenberg
Journal:  Am Heart J       Date:  1999-06       Impact factor: 4.749

10.  Cost-efficacy modeling of catheter reuse for percutaneous transluminal coronary angioplasty.

Authors:  K H Mak; M J Eisenberg; D S Eccleston; K J Brown; S G Ellis; E J Topol
Journal:  J Am Coll Cardiol       Date:  1996-07       Impact factor: 24.094

View more
  2 in total

1.  Comparison of performance characteristics between new and reprocessed electrophysiology catheters.

Authors:  Bruce R Lester; Alan A Alexander; Keith Miller; Nicole P Boser; Brian F Sullivan; Gregory G Brucker
Journal:  J Interv Card Electrophysiol       Date:  2007-01-17       Impact factor: 1.900

2.  Conception and validation of a protocol for reuse of non-irrigated electrophysiology catheters in a Brazilian teaching hospital.

Authors:  Mirtes Loeschner Leichsenring; Eliane Molina Psaltikidis; Márcio Jansen de Oliveira Figueiredo; Maria Luiza Moretti; Plínio Trabasso
Journal:  J Interv Card Electrophysiol       Date:  2017-12-11       Impact factor: 1.900

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.