Evaluation of Questor urine screening system for bacteriuria and pyuria.

M Stevens1, C J Mitchell, S A Livsey, C A MacDonald.   

Abstract

AIMS: To evaluate the Questor automated bacteriuria and pyuria screening system; to compare its performance with that of a reference method; and to assess its usefulness in a routine clinical laboratory.
METHODS: The Questor urine screening system was compared with a comprehensive regimen to detect urinary tract infection, using pour-plate viable counts to determine the numbers of bacteria present in urine samples, a wide range of other cultural methods, microscopic findings and clinical information.
RESULTS: The optimal performance in detecting significant growths was a sensitivity of 93%, a specificity of 74%, a positive predictive value of 43% and a negative predictive value of 98%. The list price per test is 0.17 pounds and the capital cost of the system is 39,950 pounds. Questor can test 50 samples an hour and can be operated by one member of the laboratory staff, who is not required to make interpretative judgments--for example, a medical laboratory assistant.
CONCLUSIONS: The sensitivity and specificity of the Questor was better than that obtained from other screening systems using the same protocol. The system was easy to use and is a useful addition to the methods available for screening for bacteriuria.

Entities:  

Mesh:

Year:  1993        PMID: 8227430      PMCID: PMC501515          DOI: 10.1136/jcp.46.9.817

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  3 in total

1.  Pyelonephritis and bacteriuria. A major problem in preventive medicine.

Authors:  E H KASS
Journal:  Ann Intern Med       Date:  1962-01       Impact factor: 25.391

2.  Analyses of the FlashTrack DNA probe and UTIscreen bioluminescence tests for bacteriuria.

Authors:  C Koenig; L J Tick; B A Hanna
Journal:  J Clin Microbiol       Date:  1992-02       Impact factor: 5.948

Review 3.  Screening urines for bacteriuria.

Authors:  M Stevens
Journal:  Med Lab Sci       Date:  1989-07
  3 in total
  1 in total

1.  Evaluation of the IRIS 939 UDx flow microscope as a screening system for urinary tract infection.

Authors:  C Hughes; M J Roebuck
Journal:  J Clin Pathol       Date:  2003-11       Impact factor: 3.411

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.