Evaluation of the IRIS 939 UDx flow microscope as a screening system for urinary tract infection.

C Hughes1, M J Roebuck.   

Abstract

AIMS: To evaluate the IRIS flow microscope for the examination of urine specimens and to compare its performance with that of manual methods; in addition, to assess its usefulness as a screening system for the detection of significant bacteriuria in a routine clinical laboratory.
METHODS: The IRIS system was compared with manual quantitative microscopy for the detection of formed elements and with a variety of culture methods to detect urinary tract infection.
RESULTS: The IRIS had a higher sensitivity for the detection of clinically significant formed elements than did manual methods, and using a combination of bacterial and white cell counts produced a sensitivity and specificity of 94.9% and 54.4%, respectively, for predicting positive significant growth, with a 98.5% predictive value for a negative result. Approximately 50% of specimens could be reported as negative on the day of receipt using the screening algorithm.
CONCLUSIONS: The sensitivity and specificity of the IRIS is equivalent to that of other automated screening systems and would remove the need for manual microscopy for most specimens. The system was easy to operate and is a useful addition to the methods currently available for the examination of urine specimens.

Entities:  

Mesh:

Year:  2003        PMID: 14600130      PMCID: PMC1770118          DOI: 10.1136/jcp.56.11.844

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  4 in total

Review 1.  Screening urines for bacteriuria.

Authors:  M Stevens
Journal:  Med Lab Sci       Date:  1989-07

2.  Evaluation of the Yellow IRIS. An automated method for urinalysis.

Authors:  C E Roe; D A Carlson; R W Daigneault; B E Statland
Journal:  Am J Clin Pathol       Date:  1986-11       Impact factor: 2.493

3.  Evaluation of Questor urine screening system for bacteriuria and pyuria.

Authors:  M Stevens; C J Mitchell; S A Livsey; C A MacDonald
Journal:  J Clin Pathol       Date:  1993-09       Impact factor: 3.411

4.  METHOD FOR THE DETECTION OF SIGNIFICANT BACTERIURIA IN LARGE GROUPS OF PATIENTS.

Authors:  D A LEIGH; J D WILLIAMS
Journal:  J Clin Pathol       Date:  1964-09       Impact factor: 3.411

  4 in total
  1 in total

1.  Bacteriuria screening by automated whole-field-image-based microscopy reduces the number of necessary urine cultures.

Authors:  Rosanna Falbo; M Roberta Sala; Silvia Signorelli; Natascia Venturi; Stefano Signorini; Paolo Brambilla
Journal:  J Clin Microbiol       Date:  2012-01-11       Impact factor: 5.948

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.