Literature DB >> 8197948

Use of a Doppler guide wire for intravascular blood flow measurements: a validation study for potential neurologic endovascular applications.

J C Chaloupka1, F Viñuela, C Kimme-Smith, J Robert, G R Duckwiler.   

Abstract

PURPOSE: To conduct a validation study of a Doppler guide wire for potential neuroendovascular applications.
METHODS: A 12-MHz, 0.018-inch Doppler guide wire was evaluated in eight swine under various blood flow conditions using two types of in vivo cerebrovascular models (physiologic and arteriovenous shunting). Flow conditions were mechanically and pharmacologically altered. Doppler average peak velocity was compared with volumetric blood flow, and flow profile corrections were calculated and analyzed. Qualitative aspects of the Doppler guide wire spectra were also assessed.
RESULTS: Plots of average peak velocity versus volumetric blood flow showed excellent linear relationships (r2 > 0.94), which were maintained at high flow conditions (average peak velocity, 99 to 236 cm/sec; volumetric blood flow, 392 to 889 mL/min). Values of flow profile correction varied from 0.43 to 0.94 and showed no consistent relationship to changes in volumetric blood flow.
CONCLUSIONS: The excellent correlation between average peak velocity and volumetric blood flow over a wide range of blood flow conditions and the additional qualitative information of the Doppler guide wire spectra establish a foundation for clinical implementation. The unpredictable variations of flow profile corrections remain obstacles for calculating volumetric blood flow based on Doppler guide wire average peak velocity.

Entities:  

Mesh:

Year:  1994        PMID: 8197948      PMCID: PMC8334306     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  6 in total

Review 1.  Interventional neuroradiology.

Authors:  C M Strother
Journal:  AJNR Am J Neuroradiol       Date:  2000-01       Impact factor: 3.825

2.  Immediate evaluation of angioplasty and stenting results in supra-aortic arteries by use of a Doppler-tipped guidewire.

Authors:  Nina Ghosh; Donatella Tampieri; Denis Melancon
Journal:  AJNR Am J Neuroradiol       Date:  2004-08       Impact factor: 3.825

3.  Therapeutic management of cerebral arteriovenous malformations. Present role of interventional neuroradiology.

Authors:  F Viñuela; G Duckwiler; R Jahan; Y Murayama
Journal:  Interv Neuroradiol       Date:  2005-10-27       Impact factor: 1.610

4.  Cerebral aneurysms treated with flow-diverting stents: computational models with intravascular blood flow measurements.

Authors:  M R Levitt; P M McGah; A Aliseda; P D Mourad; J D Nerva; S S Vaidya; R P Morton; B V Ghodke; L J Kim
Journal:  AJNR Am J Neuroradiol       Date:  2013-07-18       Impact factor: 3.825

5.  In vitro validation of endovascular Doppler-derived flow rates in models of the cerebral circulation.

Authors:  P M McGah; J D Nerva; R P Morton; M C Barbour; M R Levitt; P D Mourad; L J Kim; A Aliseda
Journal:  Physiol Meas       Date:  2015-10-09       Impact factor: 2.833

6.  Accuracy of computational cerebral aneurysm hemodynamics using patient-specific endovascular measurements.

Authors:  Patrick M McGah; Michael R Levitt; Michael C Barbour; Ryan P Morton; John D Nerva; Pierre D Mourad; Basavaraj V Ghodke; Danial K Hallam; Laligam N Sekhar; Louis J Kim; Alberto Aliseda
Journal:  Ann Biomed Eng       Date:  2013-10-26       Impact factor: 3.934

  6 in total