Literature DB >> 8194236

Failure of a polyethylene total knee component presenting as a thigh mass. Report of a rare complication of total knee arthroplasty.

D V Chavda1, K L Garvin.   

Abstract

Early failure caused by wear of the polyethylene bearing surface of a metal-backed tibial platform or a metal-backed patellar component in a total knee prosthesis has been a recognized complication; patients with these problems initially have pain and instability. A patient who had a cystic mass in the distal thigh was evaluated, and osteolysis of the distal femur was detected. Polyethylene debris and reactive synovium were confirmed surgically and histologically, indicating a cause-effect relationship of polyethylene wear debris and failure of the total knee arthroplasty. Regular and long-term patient follow-up evaluation after total joint arthroplasty is essential for the detection of early evidence of occult polyethylene wear.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8194236

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  4 in total

1.  A combined XPS-SEM/EDX investigation on explanted UHMW polyethylene acetabular cups: possible role of silicon traces in the wear debris.

Authors:  E De Giglio; A Motta; L Quagliarella; L Sabbatini; G Solarino; P G Zambonin
Journal:  J Mater Sci Mater Med       Date:  2001-01       Impact factor: 3.896

2.  Metal wear-induced pseudotumour following an endoprosthetic knee replacement for Ewing sarcoma.

Authors:  Richard Craig; Marianna Vlychou; Catherine L McCarthy; Christopher L M H Gibbons; N A Athanasou
Journal:  Skeletal Radiol       Date:  2017-03-07       Impact factor: 2.199

3.  Wear debris pseudotumor following total knee arthroplasty: a case report.

Authors:  Andreas F Mavrogenis; George N Nomikos; Vasileios I Sakellariou; George I Karaliotas; Panayiotis Kontovazenitis; Panayiotis J Papagelopoulos
Journal:  J Med Case Rep       Date:  2009-11-29

4.  A Rare Case of Pseudotumor Formation following Total Knee Arthroplasty.

Authors:  S Sivananthan; R Pirapat; S B Goodman
Journal:  Malays Orthop J       Date:  2015-03
  4 in total