| Literature DB >> 8182319 |
Abstract
This paper compares the physical parameters of the newer Groningen low resistance and Provox indwelling laryngectomy prostheses with the established and original Groningen device. In vitro pressure/flow profiles were determined, using specially designed apparatus, in 44 standard Groningen high resistance (GHR), 37 Groningen low resistance (GLR) and 19 Provox tracheo-oesophageal prostheses prior to insertion. GHR valves had significantly higher forward opening pressures than both the newer valves and the GHR was significantly higher than the Provox (p < 0.01: Mann-Whitney U-test). The mean forward resistance of GHR was significantly higher than that of both; the Provox valve was significantly lower than that of GLR (p < 0.0001: Mann-Whitney U-test). This may be of relevance with respect to patient acceptability, voice quality and effective duration of valve action.Entities:
Mesh:
Year: 1994 PMID: 8182319 DOI: 10.1017/s0022215100126659
Source DB: PubMed Journal: J Laryngol Otol ISSN: 0022-2151 Impact factor: 1.469