Literature DB >> 8165520

[The modification of programmable pacemakers by therapeutic irradiation].

M Wilm1, H L Kronholz, J Schütz, T Koch.   

Abstract

BACKGROUND: More than 300,000 pacemakers are implanted worldwide. During radiation therapy a damage of the pacemaker electronic is possible.
METHODS: Twenty pacemakers have been irradiated with photons or electrons experimentally in three different situations: a) pacemaker and pacemaker electrode outside of the irradiation field: b) pacemaker outside, pacemaker electrode inside the irradiation field; c) all things inside the irradiation field.
RESULTS: The voltage in the pacemaker electrode produced by the electric field of the accelerator did not exceed 0.8 mV if the electrode was outside the irradiation field. Induced voltage was up to 1.2 mV during irradiation with electrons (18 MeV) and the electrode being inside the treatment field with more than two thirds of its length. After delivering of not more than 10 Gy (photons) to the pacemaker, a decreasing amplitude of the pacemaker pulse occurred. The pulse frequency did not show any deviation. This seems to signal a severe early irreversible damage of the pacemaker that may cause sudden breakdown days or weeks after radiation. Two pacemakers showed a complete breakdown after irradiation with not more than 10 Gy. The others had a complete breakdown beyond doses of 50 Gy.
CONCLUSIONS: It is recommended to keep CMOS pacemakers outside the treatment field and to explant if the dose to it was higher than 10 Gy.

Mesh:

Year:  1994        PMID: 8165520

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  4 in total

Review 1.  [Radiation therapy in patients with cardiac pacemakers or implantable cardioverter defibrillators. Interdisciplinary safety recommendations].

Authors:  M Dorenkamp; C Stromberger; C von Heymann; W Haverkamp; P Wust; M Roser
Journal:  Strahlenther Onkol       Date:  2012-11-11       Impact factor: 3.621

2.  DEGRO/DGK guideline for radiotherapy in patients with cardiac implantable electronic devices.

Authors:  Benjamin Gauter-Fleckenstein; Carsten W Israel; Marc Dorenkamp; Jürgen Dunst; Mattias Roser; Rainer Schimpf; Volker Steil; Jörg Schäfer; Ulrike Höller; Frederik Wenz
Journal:  Strahlenther Onkol       Date:  2015-03-05       Impact factor: 3.621

3.  A case of radiotherapy for an advanced bronchial carcinoma patient with implanted cardiac rhythm machines as well as heart assist device.

Authors:  Sergiu Scobioala; Iris Ernst; Christos Moustakis; Uwe Haverkamp; Sven Martens; Hans Theodor Eich
Journal:  Radiat Oncol       Date:  2015-04-07       Impact factor: 3.481

4.  Management of radiation oncology patients with a pacemaker or ICD: a new comprehensive practical guideline in The Netherlands. Dutch Society of Radiotherapy and Oncology (NVRO).

Authors:  Coen W Hurkmans; Joost L Knegjens; Bing S Oei; Ad J J Maas; G J Uiterwaal; Arnoud J van der Borden; Marleen M J Ploegmakers; Lieselot van Erven
Journal:  Radiat Oncol       Date:  2012-11-24       Impact factor: 3.481