Metallosis caused by alumina ceramic screw fixation of grafted bone in a bipolar hip endoprosthesis.

M Watanabe1, H Okumura, Y Kihara, T Shibata.   

Abstract

We report a case of metallosis due to ceramic fragments in the inner bearing surface of the polyethylene cup of a bipolar hip prosthesis with fracture of the alumina ceramic screws used for grafted bone fixation. After bone-grafting acetabuloplasty with a Bateman bipolar endoprosthesis, gradual migration of the outer head occurred, and fracture of the screws was observed. Seven years after the operation, radiographs showed severe metallosis of the surrounding tissues. At revision, marked wear of the inner head was seen. Detailed examination revealed that many rough-edged ceramic particles originating from the fractured ceramic screws had become embedded in the inner surface of the polyethylene bearing insert. These extremely hard ceramic fragments had entered the prosthetic joint space during joint movement, thereafter causing continuous abrasion of the metal head. After revision, the concentration of the constituents of the prosthesis alloy in the serum decreased toward control levels. To prevent serious complications like those in the present case, screws should be inserted more than 1 cm from the outer head of a bipolar endoprosthesis, and the outer head should be removed as soon as possible if it comes into contact with the screws.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8123384     DOI: 10.1007/bf00452969

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  2 in total

1.  Massive wear of a steel ball head by ceramic fragments in the polyethylene acetabular cup after revision of a total hip prosthesis with fractured ceramic ball.

Authors:  I Kempf; M Semlitsch
Journal:  Arch Orthop Trauma Surg       Date:  1990       Impact factor: 3.067

2.  Femoral head migration after single assembly total hip arthroplasty.

Authors:  H U Cameron; P Hood-Szivek; R Turner
Journal:  Clin Orthop Relat Res       Date:  1982-04       Impact factor: 4.176

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.