Literature DB >> 8122544

Clinical evaluation of a new epidural pressure monitor.

T Czech1, A Korn, A Reinprecht, W Schramm, T Kimla, C K Spiss.   

Abstract

Comparative measuring of epidural pressure using the Spiegelberg probe 1 and ventricular fluid pressure was carried out in 15 neurosurgical intensive-care patients. Deviations in both directions were established, with a trend toward overestimating ventricular pressure in epidural pressure measuring (r = 0.77). Individual pulsations, spontaneous wave courses and therapy-induced pressure changes were reflected without delays. There were no complications observed with probe implantation periods for up to 6 days. The system has been shown to be mechanically stable and easy to implant. We believe the device to be fit for trend monitoring of intracranial pressure. As with other epidural pressure monitoring systems, false assessments of ventricular pressure may lead to wrong decisions as to required therapy.

Entities:  

Mesh:

Year:  1993        PMID: 8122544     DOI: 10.1007/bf01401846

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  6 in total

1.  A practical technique for monitoring extradural pressure.

Authors:  N W Dorsch; L Symon
Journal:  J Neurosurg       Date:  1975-03       Impact factor: 5.115

2.  Clinical evaluation of a simple epidural pressure sensor.

Authors:  M Kosteljanetz; S E Børgesen; P Stjernholm; L Christensen; O Osgaard; F Gjerris; J Rosenørn
Journal:  Acta Neurochir (Wien)       Date:  1986       Impact factor: 2.216

3.  Intracranial pressure measurements and rheological behavior of the intracranial system in dogs.

Authors:  A Schettini; J M Mahig
Journal:  J Neurosurg       Date:  1972-08       Impact factor: 5.115

4.  Behavior of an extradural pressure monitor in clinical use. Comparison of extradural with intraventricular pressure in patients with acute and chronically raised intracranial pressure.

Authors:  M P Powell; H A Crockard
Journal:  J Neurosurg       Date:  1985-11       Impact factor: 5.115

Review 5.  ICP monitoring.

Authors:  M R Gaab; H E Heissler
Journal:  Crit Rev Biomed Eng       Date:  1984

6.  Some physical principles of intracranial pressure measurement.

Authors:  G Martin
Journal:  Neurochirurgia (Stuttg)       Date:  1979-07
  6 in total
  1 in total

Review 1.  Advances in Intracranial Pressure Monitoring and Its Significance in Managing Traumatic Brain Injury.

Authors:  Usmah Kawoos; Richard M McCarron; Charles R Auker; Mikulas Chavko
Journal:  Int J Mol Sci       Date:  2015-12-04       Impact factor: 5.923

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.