| Literature DB >> 8118992 |
R B Bourne1, C H Rorabeck, B C Burkart, P G Kirk.
Abstract
Plasma spray coatings on titanium alloy total hip replacements (THRs) provide a viable alternative to sintered bead or diffusion-bonded fiber metal ingrowth surfaces. The plasma spray process enables the titanium alloy implant to retain 90% of its fatigue strength, as opposed to less than 50% for sintered or diffusion-bonded cementless devices. Laboratory and clinical evidence suggest that plasma spray cementless THRs perform as well, if not better, than other cementless implants. Clinical trials have demonstrated the superiority of a plasma spray cementless THR (Mallory-Head) over a sintered bead cementless THR (PCA) and the equivalent results to a cemented THR. Plasma spray coatings on titanium alloy hip replacements provide a safe, predictable coating for cementless fixation of THRs in the short to medium range of follow-up periods.Entities:
Mesh:
Substances:
Year: 1994 PMID: 8118992
Source DB: PubMed Journal: Clin Orthop Relat Res ISSN: 0009-921X Impact factor: 4.176