Clinical assessment of a continuous intraarterial blood gas monitoring system.

T Uchida1, K Makita, Y Tsunoda, H Toyooka, K Amaha.   

Abstract

We evaluated the clinical performance of a continuous intraarterial blood gas monitoring (CIABG) system which includes a fluorometric intravascular sensor. Seventeen patients undergoing elective surgery were monitored perioperatively with the CIABG system (PB3300; Puritan Bennett, Carlsbad, CA). Conventional laboratory blood gas analyses (BGA) were performed simultaneously whenever indicated, and the values were compared with those obtained from the CIABG system. Monitoring time ranged from 24 to 72 hr. The biases (average error between PB3300 and BGA) of pH, PCO2 and PO2 were 0.003 pH unit, -2.8 mmHg, and 0.9 mmHg in the operating room (OR), and 0.005 pH unit, 3.9 mmHg, and 8.5 mmHg in the intensive care unit (ICU), respectively. The precision (standard deviation of the bias) of pH, PCO2 and PO2 were 0.030 pH unit, 2.1 mmHg, and 29.9 mmHg in the OR and 0.035 pH unit, 3.8 mmHg, and 14.7 mmHg in the ICU, respectively. Although the PB3300 system was clinically useful as a trend monitor, the system's precision and reliability were unacceptable for estimation of true blood gas values.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8111948     DOI: 10.1007/BF03009665

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  7 in total

1.  Clinical assessment of a flow-through fluorometric blood gas monitor.

Authors:  G Bashein; J A Pino; M L Nessly; M A Kenny; K B Davis; T F Hornbein; T D Ivey
Journal:  J Clin Monit       Date:  1988-07

2.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

3.  Preliminary evaluation of an intra-arterial blood gas system in dogs and humans.

Authors:  B A Shapiro; R D Cane; C M Chomka; L E Bandala; W T Peruzzi
Journal:  Crit Care Med       Date:  1989-05       Impact factor: 7.598

4.  Progress in the development of a fluorescent intravascular blood gas system in man.

Authors:  C K Mahutte; C S Sassoon; J R Muro; D R Hansmann; T P Maxwell; W W Miller; M Yafuso
Journal:  J Clin Monit       Date:  1990-04

5.  Initial evaluation of a new intra-arterial blood gas system in humans.

Authors:  J L Zimmerman; R P Dellinger
Journal:  Crit Care Med       Date:  1993-04       Impact factor: 7.598

6.  Clinical performance of a blood gas monitor: a prospective, multicenter trial.

Authors:  B A Shapiro; C K Mahutte; R D Cane; I J Gilmour
Journal:  Crit Care Med       Date:  1993-04       Impact factor: 7.598

7.  Continuous measurement of intraarterial pHa, PaCO2, and PaO2 in the operating room.

Authors:  S J Barker; J Hyatt
Journal:  Anesth Analg       Date:  1991-07       Impact factor: 5.108

  7 in total
  2 in total

1.  Con: is continuous intra-arterial blood gas and pH monitoring justifiable?

Authors:  J L Hoffer; E A Norfleet
Journal:  J Clin Monit       Date:  1996-03

2.  Clinical evaluation of a multiparameter intra-arterial blood-gas sensor.

Authors:  E Abraham; T J Gallagher; S Fink
Journal:  Intensive Care Med       Date:  1996-05       Impact factor: 17.440

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.