Sampled gas need not be returned during low-flow anesthesia.

J P Bengtson1, J Bengtsson, A Bengtsson, O Stenqvist.   

Abstract

OBJECTIVE: The purpose of this investigation was to study the N2 flux between the patient and the breathing circuit, and the excess gas during N2O anesthesia with the low, fresh gas flow technique.
METHODS: Forty patients were studied. After a 6-minute high, fresh gas flow denitrogenation period, the O2 fresh gas flow was set at about 4 ml/kg/min and the N2O fresh gas flow was set to maintain an inspired O2 fraction of 0.30. The excess gas flow and N2 excretion were measured by a variant of the Douglas bag method.
RESULTS: The mean inspired N2 concentration reached a peak of 5.9% at 40 minutes. The estimated mean N2 excretion was 39 ml/min at 10 minutes, declining to 18 ml/min at 60 minutes. A calculation of N2 homeostasis during closed-circuit anesthesia based on the results of the patient study indicated that sampling for gas analysis actually reduces the gas costs if the sampled gas is scavenged instead of returned to the circle system, since intermittent flushing with high, fresh gas flow for denitrogenation is unnecessary in the former situation.
CONCLUSIONS: Regardless of the fresh gas flow used, sampled gas need not be returned during N2O anesthesia.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8106886     DOI: 10.1007/bf01618674

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  8 in total

Review 1.  Perioperative functional residual capacity.

Authors:  R W Wahba
Journal:  Can J Anaesth       Date:  1991-04       Impact factor: 5.063

2.  The rate of nitrogen elimination from the body through the lungs.

Authors:  C D STEVENS; H W RYDER
Journal:  J Aviat Med       Date:  1947-04

3.  Totally closed circuit nitrous oxide/oxygen anaesthesia.

Authors:  F Barton; J F Nunn
Journal:  Br J Anaesth       Date:  1975-03       Impact factor: 9.166

4.  Gaseous homeostasis during low-flow anaesthesia.

Authors:  J P Bengtson; H Sonander; O Stenqvist
Journal:  Acta Anaesthesiol Scand       Date:  1988-10       Impact factor: 2.105

5.  Accumulation of methane, acetone, and nitrogen in the inspired gas during closed-circuit anesthesia.

Authors:  S Morita; W Latta; K Hambro; M T Snider
Journal:  Anesth Analg       Date:  1985-03       Impact factor: 5.108

6.  Measurement of the rates of nitrous oxide uptake and nitrogen excretion in man.

Authors:  P C Beatty; B Kay; T E Healy
Journal:  Br J Anaesth       Date:  1984-03       Impact factor: 9.166

7.  O2 pressures between 0.12 and 2.5 atm abs, circulatory function, and N2 elimination.

Authors:  D Anderson; G Nagasawa; W Norfleet; A Olszowka; C Lundgren
Journal:  Undersea Biomed Res       Date:  1991-07

8.  Oxygen analyzers in anaesthesia: performance in a simulated clinical environment.

Authors:  J P Bengtson; H Sonander; O Stenqvist
Journal:  Acta Anaesthesiol Scand       Date:  1986-11       Impact factor: 2.105

  8 in total
  1 in total

1.  Influence of the reference gas of paramagnetic oxygen analyzers on nitrogen concentrations during closed-circuit anesthesia.

Authors:  J F Hendrickx; A A van Zundert; A M de Wolf
Journal:  J Clin Monit Comput       Date:  1998-08       Impact factor: 2.502

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.