| Literature DB >> 8106596 |
Y Teffera1, F P Abramson, M McLean, M Vestal.
Abstract
An isotope-selective detector for HPLC based on the particle-beam-interface and the chemical-reaction-interface mass spectrometer (CRIMS) principle is described. This paper focuses on the selective detection of deuterium-labeled analytes. The CRIMS product HD is detected as has been previously described for GC-CRIMS. The analytical performance was not affected by analyte structure or solvent composition. Deuterium detection was linear from 20 ng to 490 ng using 2H3-labeled cortisol. Based on this method, the fractional abundance of three labeled metabolites of cortisol were determined in the urine of a patient infused with tracer amounts of deuterated cortisol.Entities:
Mesh:
Substances:
Year: 1993 PMID: 8106596 DOI: 10.1016/0378-4347(93)80055-9
Source DB: PubMed Journal: J Chromatogr