Variations in sample pH and pO2 affect ExacTech meter glucose measurements.

E S Kilpatrick1, A G Rumley, E A Smith.   

Abstract

A study was conducted to assess the effect of changes in sample pH and pO2 on the Accutrend, ExacTech Companion, Glucometer II, One Touch II, and Refloflux II blood glucose meters. Venous blood was tonometered after the addition of strong acid or alkali to produce 10 samples with a pH range of 6.54 to 7.73 but with a similar mean pO2 (9.09 +/- 0.096 (SEM) kPa) and pCO2 (5.52 +/- 0.024 kPa). The ExacTech showed a > 15% deviation from the value obtained at pH 7.40 below pH 6.95 and above pH 7.85 (r = 0.69, p = 0.026). Ten further samples were made with a pO2 range of 2.0 to 33.6 kPa but mean pH (7.28 +/- 0.012) and pCO2 (7.98 +/- 0.25 kPa) kept alike. The ExacTech was accurate at pO2 = 11.5 kPa but deviated by > 15% below pO2 = 5.8 kPa and above 22.9 kPa (r = -0.95, p = 0.00002). The Accutrend and One Touch II required a pO2 change of 42 and 45 kPa to achieve the same error (r = -0.82 and r = -0.67 respectively). Thus caution is required in the interpretation of ExacTech glucose measurements in severely acidotic or hypoxic patients.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8088132     DOI: 10.1111/j.1464-5491.1994.tb00315.x

Source DB:  PubMed          Journal:  Diabet Med        ISSN: 0742-3071            Impact factor:   4.359


  4 in total

Review 1.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

2.  The hitchhiker's guide to research in clinical biochemistry.

Authors:  Eric S Kilpatrick
Journal:  Clin Biochem Rev       Date:  2010-02

Review 3.  Challenges to glycemic measurement in the perioperative and critically ill patient: a review.

Authors:  Andrew D Pitkin; Mark J Rice
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

4.  Comparison of Accu Chek Inform II point-of-care test blood glucose meter with Hexokinase Plasma method for a diabetes mellitus population during surgery under general anesthesia.

Authors:  B Lascaris; H G Freling; M A Edens; M J Fokkert; C G Olthof; R J Slingerland
Journal:  J Clin Monit Comput       Date:  2021-01-26       Impact factor: 1.977

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.