| Literature DB >> 8087274 |
R L Erickson1, L I Thulin, G J Richard.
Abstract
The purpose of the study was to measure the valve sound loudness of eight 25 mm aortic mechanical heart valve prostheses under controlled laboratory conditions in an isolated acoustic environment. A specially designed single-chamber pulse duplicator was constructed within an anechoic box that isolated the sounds of the valve from external noise. The valves were closed with a stimulated aortic pressure waveform of three different amplitudes. Valve sound loudness was measured with ISO-532/B (Zwicker method), which is based on a model of loudness perception that is better than dB(A), the traditional A-weighted sound pressure level. The average loudness for each valve was St. Jude Medical, 0.97 sones; CarboMedics, 0.98 sones; Sorin Biomedica Bicarbon, 1.08 sones; Edwards Tekna, 1.92 sones; Edwards Duromedics, 2.36 sones; Monostrut, 2.48 sones; Sorin Biomedica Carbocast, 3.17 sones; and Medtronic Hall, 3.57 sones. There was a positive trend between valve sound loudness and aortic drive pressure for most valves in the study.Mesh:
Year: 1994 PMID: 8087274
Source DB: PubMed Journal: J Heart Valve Dis ISSN: 0966-8519