The performance of ultrasound physiotherapy machines in Lothian Region, Scotland, 1992.

S D Pye1, C Milford.   

Abstract

Eighty-five ultrasound therapy machines in use in Lothian Region, Scotland were tested for performance and their recalibration was expedited where necessary. The performance characteristics reported are the temporal average acoustic output power, the treatment frequency, pulse timing and treatment timer accuracy. Consideration is also given to the performance of the wide dynamic range radiation force balance that was used to make acoustic power measurements. Of the machines tested, 69% had power outputs that differed by more than 30% from the expected values. The results of power measurements are analysed according to continuous and pulsed output, single and dual frequency treatment heads and year of manufacture. Therapy machines more than 10-12 years old and modern dual frequency treatment heads performed particularly badly. This study suggests that the performance of ultrasound therapy machines has improved little in the last 20 years despite the introduction of IEC 601-2-5 in 1984.

Mesh:

Year:  1994        PMID: 8085291     DOI: 10.1016/0301-5629(94)90003-5

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  2 in total

1.  Development and application of a quality control procedure for short-wave diathermy units.

Authors:  N Shields; N O'Hare; G Boyle; J Gormley
Journal:  Med Biol Eng Comput       Date:  2003-01       Impact factor: 2.602

2.  Variability in effective radiating area and output power of new ultrasound transducers at 3 MHz.

Authors:  Lennart D Johns; Stephen J Straub; Samuel M Howard
Journal:  J Athl Train       Date:  2007 Jan-Mar       Impact factor: 2.860

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.