Literature DB >> 12572749

Development and application of a quality control procedure for short-wave diathermy units.

N Shields1, N O'Hare, G Boyle, J Gormley.   

Abstract

Short-wave diathermy (SWD) is a form of radiofrequency (RF) radiation, operating at 27.12 MHz, that is used therapeutically by physiotherapists. Although this form of therapy is widely available, the management of the equipment is not often addressed by either physiotherapists or by medical physics/clinical engineering. A quality control protocol for SWD units, examining power output and electrical and mechanical condition, was developed and applied to 20 units used in clinical practice. In addition, an environmental assessment of where the units were used was also included. Results showed that the power output was generally stable (coefficient of variation range 0-8.8%) and reproducible (coefficient of variation range 0-6.8%). When the outputs from 12 similar units were compared, it was found that the relationship between the units' intensity settings and power output measurements was non-linear. Two units with mechanical timers were found to have inaccuracies that could contribute, under a 'worst-case' scenario, to a dosage error of up to 45%. Environmental analysis found that all treatment plinths in use contained metal parts, which could constitute a fire hazard, and no department examined was equipped with an RF screened room, a facility that would ensure that other persons in the vicinity were not exposed to excessive stray radiation.

Entities:  

Mesh:

Year:  2003        PMID: 12572749     DOI: 10.1007/BF02343540

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  3 in total

1.  Guidelines on limits of exposure to radiofrequency electromagnetic fields in the frequency range from 100 kHz to 300 GHz. International Non-Ionizing Radiation Committee of the International Radiation Protection Association.

Authors: 
Journal:  Health Phys       Date:  1988-01       Impact factor: 1.316

2.  Evaluation report: shortwave therapy units.

Authors: 
Journal:  J Med Eng Technol       Date:  1987 Nov-Dec

3.  The performance of ultrasound physiotherapy machines in Lothian Region, Scotland, 1992.

Authors:  S D Pye; C Milford
Journal:  Ultrasound Med Biol       Date:  1994       Impact factor: 2.998

  3 in total
  2 in total

1.  ELECTROPHYSICAL AGENTS - Contraindications And Precautions: An Evidence-Based Approach To Clinical Decision Making In Physical Therapy.

Authors: 
Journal:  Physiother Can       Date:  2011-01-05       Impact factor: 1.037

2.  Lack of maintenance of shortwave diathermy equipment has a negative impact on power output.

Authors:  Rinaldo Roberto de Jesus Guirro; Elaine Caldeira de Oliveira Guirro; Natanael Teixeira Alves de Sousa
Journal:  J Phys Ther Sci       Date:  2014-04-23